Objective: To evaluate the accuracy and safety of a novel patient-specific drill guide template for stabilizing the thoracolumbar vertebrae of dogs.

Study Design: Cadaveric experimental study and prospective case series.

Sample Population: Cadaveric canine thoracolumbar vertebral specimens (n = 3) and clinical cases of thoracolumbar spinal instability (n = 4).

Methods: Computed tomography data of the thoracolumbar spines were obtained before surgery, and images were imported into imaging software. Optimum screw trajectories were selected for each vertebra, and drill guide templates were designed and fabricated with a 3-dimensional printing system. Drill guide templates were applied to cadaveric spine and clinical cases. Computed tomography imaging was performed after surgery, and planned and postoperative trajectories were compared to estimate the accuracy and safety of the drill guide templates.

Results: Twenty-two drill holes were made in cadaveric spinal specimens. All drill holes were completely located in the bone. The overall mean screw deviation was 0.88 ± 0.36 mm. In clinical cases, 29 screws were placed in thoracolumbar vertebrae. Most (89.6%) of these screws were placed without evidence of vertebral canal invasion. One (3.5%) screw perforated the bone structure. The overall mean screw deviation was 1.16 ± 0.56 mm.

Conclusion: Drill guide templates were useful for accurate intraoperative screw navigation in thoracolumbar fixation in small dogs.

Clinical Significance: The use of drill guide templates can be considered as an aid to safety and accuracy of screw placement in canine thoracolumbar instabilities.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vsu.13140DOI Listing

Publication Analysis

Top Keywords

drill guide
28
clinical cases
16
guide templates
16
thoracolumbar vertebrae
12
drill
9
novel patient-specific
8
patient-specific drill
8
guide template
8
thoracolumbar
8
accuracy safety
8

Similar Publications

Forehead osteomas are benign tumors commonly excised for cosmetic and functional reasons. However, removing osteomas from the anterior table of the frontal sinus presents specific challenges, particularly in determining the appropriate removal thickness. Inaccurate resection depth may result in fracture or perforation of the anterior table of the frontal sinus, or incomplete resection.

View Article and Find Full Text PDF

Revision quadriceps tendon repair: A case series and technique guide to a novel repair.

Trauma Case Rep

February 2025

Department of Orthopaedic Surgery and Sports Medicine, University at Buffalo, Jacobs School of Medicine and Biomedical Sciences, Buffalo, NY 14203, United States of America.

Introduction: Revision quadriceps tendon repair is a challenging problem. In this four-case series, novel quadriceps tendon revision resulted in improved range of motion and durable repair for patients with recurrent rupture.

Methods: Our technique includes a combination of a running locked #5 FiberWire or 2 mm SutureTape suture placed through parallel medial, lateral, and central drill holes in the patella with running Krackow-type quadriceps tendon repair medially and laterally resulting in four strands, delivering the vastus medialis and medial quadriceps tendon to an anatomic repair at the superior pole of the patella, with 2 sutures passed centrally and 1 each passed medially and laterally and then tied.

View Article and Find Full Text PDF

 Minimally invasive Ponto surgery (MIPS) enables the installation of percutaneous bone-anchored hearing implants (BAHIs) with a drill guide through a hole punch incision. Despite being well established for adults, there is a lack of studies in the literature regarding its use in pediatric patients.  The aim of the present study was to investigate the hearing performance and soft-tissue outcomes of the use of MIPS under local anesthesia in children with unilateral craniofacial malformation (UCM).

View Article and Find Full Text PDF

Background: In the maxillary anterior region, differences in bone density along the drilling path can impact the accuracy of implant placement despite the use of a surgical guide. Hence, the aim of this in vitro study was to investigate the accuracy of implant placement using the over-preparation technique in different drilling environments in the maxillary anterior region.

Methods: Three experimental models (a, b, and c) were designed and fabricated to simulate the following drilling environments (n = 60 each): unhealed bone, less dense bone, and dense bone after tooth extraction.

View Article and Find Full Text PDF

Purpose: The aim of this study was to evaluate the feasibility of using patient-specific implants (PSI) for complex shaft corrective osteotomies in multiplanar deformities of long bones in the lower extremities. Additionally, it aimed to investigate the added value of these implants by quantifying surgical accuracy on postoperative CT, comparing their outcomes to two commonly used techniques: 3D virtual visualizations and 3D-printed surgical guides.

Methods: Six tibial and femoral shaft corrective osteotomies were planned and performed on three Thiel embalmed human specimen.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!