Purpose: Evaluation of the long-term efficiency of MIGS implants is still challenging, due to the lack of standardized clinical studies of stand-alone procedures. Moreover, the different mechanisms of the various glaucoma drainage devices are not adequately considered. The current study focusses on the development of a method for oculopression to evaluate the efficiency of glaucoma drainage devices.
Methods: Explanted porcine eyes were subjected to pressure or weight load using three oculopressors with different modi operandi. The time-dependent intraocular pressure was measured using an anterior chamber maintainer. The Honan Balloon exerts variable pressure onto the eye via an air bellows, whereas the Taylor oculopressor applies a defined weight loading on the eye. A novel oculopressor with a weight loading of 60 g was developed and manufactured by means of 3-D-printing.
Results: The intraocular pressure changes observed during the experiments were similar for all tested oculopression devices, varying only in the absolute pressure values. The Honan Balloon was not suitable for the intended purpose, due to poor standardisation of the applied pressure. Oculopression using a defined weight appeared more suitable. The Taylor oculopressor, however, created intraocular pressure values of up to 203.3 ± 38.4 mmHg, which precludes its use with glaucoma patients. On the basis of these data, the new oculopression device was used in a preliminary trial with healthy human subjects, thereby preparing its use in a clinical trial.
Conclusions: Oculopression represents a potentially suitable tool to analyse the efficiency of glaucoma drainage devices. Commercially available oculopression devices are not directly applicable for this task. Difficult handling, high intraocular pressure, and lack of standardisation complicate the use for glaucoma patients. These difficulties were overcome with the newly designed oculopressor that facilitates a well defined increase in intraocular pressure. The device is currently being used in a clinical study to evaluate the efficiency of MIGS implants.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1055/a-0792-1247 | DOI Listing |
J Control Release
January 2025
Department of Materials Science and Engineering, Pohang University of Science and Technology (POSTECH), 77 Cheongam-ro, Nam-gu, Pohang, Gyeongbuk 37673, Republic of Korea. Electronic address:
Although smart contact lenses have demonstrated great potential in theranostics, there remain critical challenges and opportunities in their commercial development. In this Perspective, the current status and capability of smart theranostic contact lenses are highlighted, focusing on their application as sensing systems for detecting biomarkers such as glucose, intraocular pressure (IOP), and inflammatory cytokines, and as drug delivery systems (DDS) for precise and controlled therapy. Additionally, key challenges associated with clinical development and commercialization of smart theranostic contact lenses are discussed, to optimize diagnostic and therapeutic interventions.
View Article and Find Full Text PDFPLoS One
January 2025
Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Purpose: To investigate the heritability of genetic influence on macular choroidal vascularity index (CVI).
Methods: Total choroidal area (TCA), luminal area (LA), and CVI was measured using horizontal scan of spectral-domain optical coherence tomography with enhanced depth imaging in the 373 healthy twin participants. Characteristics of the participants were investigated, including age, sex, axial length, hypertension, diabetes, drinking habits, and smoking status.
Optom Vis Sci
January 2025
Department of Anatomy and Physiology, The University of Melbourne, Parkville, Victoria, Australia.
Significance: Previous evidence showed that transient receptor potential vanilloid 4 (TRPV4) inhibition was protective of retinal ganglion cell (RGC) loss after chronic intraocular pressure (IOP) elevation in young animals. However, the role of TRPV4 in mechanosensing IOP changes in the aging eye is not well understood.
Purpose: This study compared the recovery of retinal function and structure after acute IOP elevation in 3- and 12-month-old mouse eyes with and without TRPV4 inhibition.
Front Med (Lausanne)
January 2025
Eye Center of the Second Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, Zhejiang, China.
Introduction: Glaucoma, a leading cause of irreversible blindness, is characterized by optic neuropathy and retinopathy, with primary open-angle glaucoma (POAG) being the most prevalent form. The primary pathogenic mechanism of POAG involves elevated intraocular pressure caused by chronic fibrosis of the trabecular meshwork (TM). Autophagy, a critical process for maintaining cellular homeostasis, has been implicated in fibrosis across various organs.
View Article and Find Full Text PDFBMC Genom Data
January 2025
School of Epidemiology and Public Health, University of Ottawa, 600 Peter Morand Crescent, Office 101E, Ottawa, Ontario, K1G 5Z3, Canada.
High intraocular pressure (IOP) is an important risk factor for glaucoma, which is influenced by genetic and environmental factors. However, the etiology of high IOP remains uncertain. Metabolites are compounds involved in metabolism which provide a link between the internal (genetic) and external environments.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!