Dose estimation of ultra-low-dose chest CT to different sized adult patients.

Eur Radiol

Department of Organismal Biology, Uppsala University, 752 36, Uppsala, Sweden.

Published: August 2019

Objectives: To evaluate the effect of patient size on radiation dose for standard CT (SD-CT), ultra-low-dose CT (ULD-CT) and two-view digital radiography (DR).

Methods: Dosimeters were distributed within the lungs of chest phantoms representing males of 65 kg and 82 kg (body mass indices 23 and 29). In contrast to SD-CT and DR which include automatic exposure control (AEC), the ULD scan employs a fixed mAs value. The phantoms were exposed to SD, ULD and DR while recording lung doses. Projected dose data were calculated from the phantoms. The resulting exposure settings were used in Monte Carlo programs to determine the effective dose for a standard-sized (BMI 24.2) adult male (170 cm/70 kg) and female (160 cm/59 kg). Patients previously examined by both ULD- and SD-CT were identified to determine post hoc size-specific dose estimates (SSDEs).

Results: ULD-CT dose was inversely related to patient size; average lung doses summarised in terms of patient size BMI are 5.2/8.1 (SD-CT), 0.56/0.35 (ULD-CT) and 0.05/0.13 mGy (DR), while the effective doses for these techniques on a standard-sized male were 2.9, 0.16 and 0.03 mSv and 2.3, 0.247 and 0.024 mSv for a standard-sized female respectively. SSDEs for 15 patients (averages: BMI 26, range 18-37) averaged 5.5 mGy (3.6-10) for SD-CT and 0.35 mGy (0.42-0.27) for ULD-CT.

Conclusions: The effective doses for a standard-sized male and female examined by ULD-CT are (respectively) ~ 6%/~ 11% of SD-CT and ~ 5/~ 10 times higher than DR. ULD-CT gave a lower radiation dosage to larger patients than DR. AEC is warranted in ULD-CT for improved dose consistency.

Key Points: • For standard-sized patients, ULD-CT dose level is ~ 6%/~ 11% of SD-CT, and ~ 5/~ 10 times higher than DR. For larger patients, ULD-CT is currently being used clinically at lower dose levels than DR. • Using ULD-CT should greatly reduce the risk of late effects from ionising radiation. • AEC in ULD-CT is desirable for increased consistency in patient dose.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-018-5849-5DOI Listing

Publication Analysis

Top Keywords

patient size
12
dose
10
uld-ct
10
lung doses
8
uld-ct dose
8
effective doses
8
standard-sized male
8
~ 6%/~ 11% sd-ct
8
sd-ct ~ 5/~ 10
8
~ 5/~ 10 times
8

Similar Publications

Background: Diffusing alpha-emitters Radiation Therapy ("Alpha DaRT") is a promising new radiation therapy modality for treating bulky tumors. Ra-carrying sources are inserted intratumorally, producing a therapeutic alpha-dose region with a total size of a few millimeter via the diffusive motion of Ra's alpha-emitting daughters. Clinical studies of Alpha DaRT have reported 100% positive response (30%-100% shrinkage within several weeks), with post-insertion swelling in close to half of the cases.

View Article and Find Full Text PDF

Background: Tumor size (TS) in pancreatic ductal adenocarcinoma (PDAC) is one of the most important prognostic factors. However, discrepancies between TS on preoperative images (TSi) and pathological specimens (TSp) have been reported. This study aims to evaluate the factors associated with the differences between TSi and TSp.

View Article and Find Full Text PDF

Background: This study aimed to investigate the clinical characteristics and esophageal motility of patients with gastric cardia submucosal tumors (SMTs) and the associated changes after endoscopic resection based on high-resolution impedance manometry (HRIM).

Methods: From our electronic database, we identified patients who underwent pre-operative evaluation of gastric cardia SMTs between 2015 and 2023. All patients completed standardized symptom questionnaires and underwent endoscopic ultrasonography and HRIM.

View Article and Find Full Text PDF

Short-term outcomes of mesh-suture repair in the treatment of ventral hernias: a single-center study.

Surg Endosc

January 2025

Division of Minimally Invasive and Bariatric Surgery, Penn State Health Milton S. Hershey Medical Center, 500 University Drive, Hershey, PA, 17033, USA.

Background: Defect closure with mesh suture is a novel technique for hernia repair. Originally described as the construction of lightweight macroporous polypropylene mesh strips as a suture material, it is now available as an FDA-approved product. Mesh suture better distributes tensile forces and reduces fascial tearing compared to traditional suture but requires less implanted material and tissue dissection compared to planar mesh.

View Article and Find Full Text PDF

Background: The avoidable causes of infant mortality should be identified, and interventions should be made to improve the infant mortality rate. The cause of infant deaths should be assessed in both medical and social contexts.

Objectives: We aimed to determine the medical causes of infant mortality by verbal autopsy and its determinants in two rural blocks of the Khordha district of Odisha and assess the pathway of care and delay in seeking care for the illness preceding infant death using the three-delay model.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!