Ex-vivo and in vitro validation of an innovative mandibular condyle implant concept.

J Craniomaxillofac Surg

University of Bordeaux, Institut de Mécanique et d'Ingénierie, CNRS UMR 5295, FR-33405 Talence, France.

Published: January 2019

Purpose: The purpose of this study is to pre-validate an inovative implant concept, and to compare the behavior of the mandibular condyle against a commercial Biomet implant in an ex vivo model and present results of the first cadaveric studies.

Materials And Methods: Three experimental cadaveric condyles were tested under three conditions: one intact, another with the Biomet model, and one with the innovative concept. The condyle was tested with a reaction of 300 N in all situations and the principal strains were measured. Before the geometry of the cadaveric condyle was reconstructed from a microCT scan, and a finite element model was created. Finally, a procedure was carried out with the new implant by two expert surgeons on a two cadaveric head model.

Results: In vitro the mandible condyle presents a linear behavior until maximum load. The strain measured with Biomet implant indicates a strain shielding effect in the proximal region, inducing bone loss in the long term. The lingual side of the Biomet implanted condyle presents an increase of +44% in strain.

Conclusion: The new concept was evaluated and showed a similar behavior to the intact model, and better behavior than the Biomet. The innovative concept proves that it is possible to avoid screws for a TMJ fixation and improve the TMJ alloplastic behavior.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcms.2018.11.010DOI Listing

Publication Analysis

Top Keywords

mandibular condyle
8
implant concept
8
biomet implant
8
innovative concept
8
condyle presents
8
condyle
6
implant
5
concept
5
behavior
5
biomet
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!