Novel distortion correction method for diffusion-weighted imaging based on non-rigid image registration between low b value image and anatomical image.

Magn Reson Imaging

Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Japan. Electronic address:

Published: April 2019

Purpose: This study aimed to develop a novel technique for retrospective distortion correction based on non-rigid image registration in magnetic resonance diffusion image.

Methods: A 3.0 T MRI scanner with an 18-channel dedicated breast coil and the outer shell of the original breast phantom, which provided images with non-uniform fat-suppression based on clinical data were used. The diffusion-weighted imaging with and without parallel imaging (PI) was used. The proposed study included several steps, which are FOV size matching, matrix size matching, image segmentation, edge detection, non-rigid image registration, and image wrap. We compared the results obtained using the proposed method with that obtained using TOPUP images. The correlation was assessed between T1-weighted image with fat suppression (FS-T1WI) and b1000 image with the help of cross-correlation coefficient (CCC). Shape-error analysis of tumor model and apparent diffusion coefficient (ADC) was calculated. The Steel-Dwass multiple-comparison tests were used for all comparisons and statistical analysis (P < 0.05).

Results: The novel method of CCC showed the highest correlation between FS-T1WI and b1000 images. In the Steel-Dwass multiple-comparison test, significant differences were found (P < 0.05) except between non-correction and TOPUP (P = 0.99). The novel method was the lowest degree of error. With PI in the right breast, no significant differences, whereas in the left breast, significant differences were observed except for between novel method and TOPUP (P = 0.73). Without PI in the right breast, significant differences were observed. In the left breast, no significant differences were observed between any combinations. The ADC value, no significant differences were observed for non-correction and novel methods.

Conclusions: We developed a novel technique for retrospective distortion correction based on non-rigid image registration. The high degree of accuracy of this method combined with the lack of requirement for additional scans renders it a promising tool for application in clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.mri.2018.12.002DOI Listing

Publication Analysis

Top Keywords

non-rigid image
12
image registration
12
image
9
distortion correction
8
diffusion-weighted imaging
8
based non-rigid
8
size matching
8
novel distortion
4
correction method
4
method diffusion-weighted
4

Similar Publications

Purpose: In laparoscopic liver surgery, registering preoperative CT-extracted 3D models with intraoperative laparoscopic video reconstructions of the liver surface can help surgeons predict critical liver anatomy. However, the registration process is challenged by non-rigid deformation of the organ due to intraoperative pneumoperitoneum pressure, partial visibility of the liver surface, and surface reconstruction noise.

Methods: First, we learn point-by-point descriptors and encode location information to alleviate the limitations of descriptors in location perception.

View Article and Find Full Text PDF

One of the main limitations of conventional absorption-based X-ray micro-computed tomography imaging of biological samples is the low inherent X-ray contrast of soft tissue. To overcome this limitation, the use of ethanol as contrast agent has been proposed to enhance image contrast of soft tissues through dehydration. Some authors have shown that ethanol shrinks and hardens the tissue too much, also causing small tissue ruptures due to fast dehydration.

View Article and Find Full Text PDF

: Prostate-specific membrane antigen positron emission tomography/computed tomography (PSMA PET/CT), in combination with magnetic resonance imaging (MRI), may enhance the diagnosis and staging of prostate cancer. Image fusion of separately acquired PET/CT and MRI images serve to facilitate clinical integration and treatment planning. This study aimed to investigate different PSMA PET/CT and MRI image fusion workflows for prostate cancer visualisation.

View Article and Find Full Text PDF

Improved myocardial scar visualization using free-breathing motion-corrected wideband black-blood late gadolinium enhancement imaging in patients with implantable cardiac device.

Diagn Interv Imaging

December 2024

IHU LIRYC, Heart Rhythm Disease Institute, Université de Bordeaux, INSERM U1045, 33604, Pessac, France; Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, 1011 Lausanne, Switzerland; Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, 33604 Pessac, France. Electronic address:

Purpose: The purpose of this study was to introduce and evaluate a novel two-dimensional wideband black-blood (BB) LGE sequence, incorporating wideband inversion recovery, wideband T2 preparation, and non-rigid motion correction (MOCO) reconstruction, to improve myocardial scar detection and address artifacts associated with implantable cardioverter defibrillators (ICDs).

Materials And Methods: The wideband MOCO free-breathing BB-LGE sequence was tested on a sheep with ischemic scar and in 22 patients with cardiac disease, including 15 with cardiac implants, at 1.5 T.

View Article and Find Full Text PDF

Quantifying spatiotemporal dynamics during embryogenesis is crucial for understanding congenital diseases. We developed Spateo (https://github.com/aristoteleo/spateo-release), a 3D spatiotemporal modeling framework, and applied it to a 3D mouse embryogenesis atlas at E9.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!