An evaluation of computed tomography dose index measurements using a pencil ionisation chamber and small detectors.

J Radiol Prot

Department of Physics, Faculty of Mathematics and Natural Sciences, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia.

Published: March 2019

The aim of this study was to compare the values of the computed tomography dose index 100 (CTDI) obtained using two small detectors (i.e. a small ionisation chamber and a small solid state detector) with those obtained from a 100 mm pencil ionisation chamber for various input CT parameters: beam width, kVp, mAs, pitch, and head-body phantom variation. The measurement of CTDI using the 100 mm pencil chamber was carried out in a single rotation of axial mode, while the measurement using small detectors was carried out in helical mode. The differences of CTDI values obtained with two small detectors were about 7% for all variations. The differences of CTDI values obtained with small detectors and a 100 mm pencil ionisation chamber for beam widths of more than 4 mm were within 40%. However, for the narrowest beam widths (4 mm), the difference between them was very large (about 150%).

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6498/aaf2b4DOI Listing

Publication Analysis

Top Keywords

small detectors
20
ionisation chamber
16
pencil ionisation
12
100 pencil
12
computed tomography
8
tomography dose
8
chamber small
8
differences ctdi
8
ctdi values
8
values small
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!