  • A study evaluated the impact of a deep learning denoising algorithm, PixelShine (PS), on the quality of 70 kVp pelvic CT images from 33 patients with a low body mass index.
  • The results showed that images processed with PS (group B) had significantly better subjective quality scores, lower noise levels, and improved signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) compared to the original images (group A).
  • The findings suggest that using PS enhances the quality of pelvic arterial phase CT images by significantly reducing noise and boosting SNR and CNR.

Article Abstract

Objective: To investigate the effect of a deep learning-based denoising algorithm, PixelShine (PS), on the quality of 70 kVp pelvic arterial phase CT images.

Materials And Methods: A retrospective analysis was performed on arterial phase pelvic CT images from 33 patients (body-mass index ≤ 20 kg/m) obtained with a GE Revolution CT (70 kVp tube voltage; adaptive statistical iterative reconstruction-Veo-filtered back projection, 50% blending) and designated group A. Group B images were then obtained by applying PS to group A image datasets. Subjective image quality was evaluated by two radiologists with a 5-point scoring system; the scores of the groups were compared. Image signal was assessed using CT values of the urinary bladder. CT and standard deviation (SD) values of the gluteus maximus were measured, and SD values of the gluteus maximus were used to represent image noise. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the bladder were calculated. Image noise, SNR, and CNR of two groups were compared using paired t-tests.

Results: The subjective visual image quality scores of groups A and B, respectively, were 3.11 ± 0.30 vs. 3.82 ± 0.57; image noise was 15.79 ± 2.05 Hounsfield units (HU) vs. 11.06 ± 2.22 HU; SNRs of bladder were 0.50 ± 0.23 vs. 0.79 ± 0.39; and CNRs of bladder were 3.72 ± 0.85 vs. 5.14 ± 1.27. Group B showed better subjective image quality, lower image noise, and improved SNR and CNR, compared to group A; these differences were statistically significant (P < 0.05). The noise of group B was approximately 30% lower than that of group A; the SNR and CNR values of group B were improved by approximately 58% and 38%, respectively.

Conclusion: Using 70 kVp +ASiR-V, PS can improve the image quality of pelvic arterial phase CT images, significantly reduce the image noise, and improve the SNR and CNR.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11604-018-0798-0DOI Listing

Publication Analysis

Top Keywords

image noise
16
arterial phase
12
image quality
12
image
9
pelvic arterial
8
subjective image
8
scores groups
8
groups compared
8
values gluteus
8
gluteus maximus
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!