Introduction: Due to the high noise emission generated by the gradients in magnetic resonance imaging (MRI), an efficient method of noise protection is mandatory. In addition to providing hearing protection, appropriate headphone systems also serve to facilitate communication between the operator and the patient. However, in combined PET-MR devices, use of common pneumatic headphones, as delivered by the manufacturer, is problematic due to the potential generation of attenuation artefacts in the PET measurement. Furthermore, modern multichannel head coils rarely provide space for conventional headphones. This work presents an alternative system, which aims to address these limitations while still being appropriate for both patient noise protection and communication in PET-MR.

Material And Methods: As an alternative to the standard headphones supplied with the PET-MR (3T MR-BrainPET, Siemens), the possibility of using earphones built out of commercially available earplugs has been investigated. The air channel (E-A-RLink) of the earplug is connected to the tubes of the original headphones. The attenuation characteristics of the conventional headphones and of the modified earphones were measured using a dedicated PET system with a Ge transmission source. For this purpose, the headphones, and then the earphones, were attached to a non-radioactive head phantom. To investigate the influence of the different phones on PET emission images, measurements of the head phantom, filled with F solution, were performed in the PET-MR. A measurement of the head phantom without headphones or earphones was used as a reference.

Results: The linear attenuation coefficient of the headphones was 0.11 cm and that of the head phantom 0.10 cm. The earphones were not identifiable in the transmission image. The emission image showed an activity underestimation of 10% near the headphones, compared to the reference image, whereas the earphones did not affect the image. Communication with the patient via the earphones was successful, and the noise protection-as confirmed by investigated subjects-was satisfying.

Conclusion: The presented earphones, which can be connected to the existing patient communication system, are a preferable alternative to the conventional headphones, as, in contrast to the use of headphones, qualitative and quantitative errors in the PET images can be avoided. Patient acceptance of the earphones was high, despite the increase in preparation time before the PET-MR study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6277401PMC
http://dx.doi.org/10.1186/s13550-018-0457-6DOI Listing

Publication Analysis

Top Keywords

head phantom
16
noise protection
12
conventional headphones
12
headphones
11
earphones
9
patient noise
8
protection communication
8
headphones earphones
8
patient
6
noise
5

Similar Publications

: This study aims to evaluate the impact of various weighting factors (WFs) on the quality of weighted average (WA) dual-energy computed tomography (DECT) non-contrast brain images and to determine the optimal WF value. Because they simulate standard CT images, 0.4-WA reconstructions are routinely used.

View Article and Find Full Text PDF

The Influence of Posterior Class II Composite Restoration Location and Techniques on Marginal Sealing.

Dent J (Basel)

January 2025

Department of Oral Rehabilitation, The Maurice and Gabriela Goldschleger School of Dental Medicine, Faculty of Medicine, Tel Aviv University, Tel Aviv 6997801, Israel.

: The success of treatment and prevention for secondary caries hinges significantly on the techniques employed in Class II composite restoration. Additionally, the location of the restored tooth within the oral cavity has emerged as a potential factor determining the quality of the restoration. A comprehensive understanding of these interrelated variables is crucial for advancing the efficacy and durability of dental composite restorations.

View Article and Find Full Text PDF

Heterogeneous head phantom for validating treatment planning system in boron neutron capture therapy.

Appl Radiat Isot

January 2025

Institute of Nuclear Engineering and Science, National Tsing Hua University, 101, Sec. 2, Kuang-Fu Road, Hsinchu, 30013, Taiwan; Nuclear Science and Technology Development Center, National Tsing Hua University, 101, Sec. 2, Kuang-Fu Road, Hsinchu, 30013, Taiwan. Electronic address:

In clinical boron neutron capture therapy (BNCT), the distribution of dose to a heterogeneous medium that is predicted by a treatment planning system (TPS) should be experimentally validated. A head phantom specifically developed for this purpose is described and demonstrated herein. The cylindrical phantom exhibits distinct regions made from four materials (polymethyl methacrylate, calcium phosphate, air, and boric acid) to approximate a head structure with explicitly defined skin, skull, and brain tissue with a cavity and tumor within.

View Article and Find Full Text PDF

Significance: In the last years, time-resolved near-infrared spectroscopy (TD-NIRS) has gained increasing interest as a tool for studying tissue spectroscopy with commercial devices. Although it provides much more information than its continuous wave counterpart, accurate models interpreting the measured raw data in real time are still lacking.

Aim: We introduce an analytical model that can be integrated and used in TD-NIRS data processing software and toolkits in real time.

View Article and Find Full Text PDF

Background: Ultrasound education is transitioning from in-person training to remote methods using mixed reality (MR) and 5G networks. Previous studies are mainly experimental, lacking randomized controlled trials in direct training scenarios.

Objective: This study aimed to compare an MR-based telesupervised ultrasound education platform on private 5G networks with traditional in-person training for novice doctors.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!