Background: MRI is predictive of adverse local tissue reactions (ALTRs) after THA but how MRI directly relates to implant surface wear, fretting, and trunnion corrosion at different articulations between implant components remains unclear. MRI generates high-contrast images to display soft tissues around arthroplasty and may provide a surgeon the means to distinguish and differentiate host-related synovial patterns as a response to either polyethylene wear or metal wear and corrosion products.
Questions/purposes: The purposes of this study were (1) to correlate findings from MRI in patients who have undergone THA with direct assessment of implant wear, corrosion, and fretting from retrieved components; and (2) to distinguish the unique synovial responses on MRI in patients who have undergone THA based on bearing materials.
Methods: In this prospective study, patients undergoing THA (181 patients, 187 hips) with metal-on-metal (MoM), hip resurfacing (HRA), metal-on-polyethylene (MoP), ceramic-on-polyethylene, ceramic-on-ceramic, or modular neck designs having revision surgery (between October 2013 and June 2017) underwent preoperative MRI. A single reader blinded to the bearing surface made an assessment of the synovial response (Gwet's AC1, 0.65-0.97); these data were compared with semiquantitative histology of tissue samples by a single reader (Gwet's AC1, 0.92) and semiquantitative wear, corrosion, and fretting analysis of retrieved components using Goldberg scoring (Gwet's AC1, 0.60-0.79). Direct noncontact measurements of implant wear were also made. Correlations and analyses of variance were used to assess associations between metrics and differences by implant type, respectively.
Results: Correlations were found between MRI synovial thickness with severity of fretting and corrosion damage of the female head-neck trunnion of femoral stems in modular designs (ρ = 0.26 [95% confidence interval {CI}, 0.12-0.39]; p = 0.015, n = 185) and ALTR grade and volumetric wear in MoM bearings (ρ = 0.93 [95% CI, 0.72-0.98]; p < 0.001, n = 10). MRI synovial thickness was highest in patients identified with aseptic lymphocyte-dominated vasculitis-associated lesions and diffuse tissue necrosis. On MRI, MoP hips demonstrated a distinct polymeric synovial response, whereas HRA, MoM, and modular hips more commonly demonstrated ALTR. Hips classified as having a polymeric synovial response on MRI had a greater number of particles present in tissue samples.
Conclusions: In this study, we demonstrated that MRI of THA can distinguish synovial responses that reflect the bearing type of the implanted THA and correlate to direct measurements of implant wear, corrosion, and fretting and histologic assessment of wear particles in periprosthetic tissues. MRI provides a means of direct, noninvasive visualization of the host-generated synovial response. Patients presenting with painful arthroplasties may be evaluated for the cause of their discomfort, specifically highlighting any concerning synovial reactions that would warrant more prompt surgical intervention. Future studies would benefit from a prospective evaluation of different implants to assess the natural longitudinal history of arthroplasty complications, including the development and prevalence of ALTR across bearing constructs.
Level Of Evidence: Level III, diagnostic study.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6345304 | PMC |
http://dx.doi.org/10.1097/CORR.0000000000000535 | DOI Listing |
Commun Med (Lond)
January 2025
MED-EL Elektromedizinische Geräte GmbH, Fürstenweg 77a, 6020, Innsbruck, Austria.
Background: Cochlear implants (CIs) are neuroprosthetic devices which restore hearing in severe-to-profound hearing loss through electrical stimulation of the auditory nerve. Current CIs use an externally worn audio processor. A long-term goal in the field has been to develop a device in which all components are contained within a single implant.
View Article and Find Full Text PDFFirst metatarsophalangeal (MTP) joint fusion is a frequently employed surgical treatment option for hallux rigidus and hallux valgus. Implant-related complications are common, necessitating further investigation into predisposing factors. The altered mechanics of pes planus may influence surgical outcomes; however, its direct impact on implant removal rates post-fusion remains unclear.
View Article and Find Full Text PDFJ Knee Surg
January 2025
Orthopaedic Surgery, Mayo Clinic, Rochester, United States.
Introduction: Historically, cementless patellar implants were fraught with issues of fixation and polyethylene wear. However, contemporary cementless patellar implants incorporate modern technologies for implant design and offer the potential for improved biological fixation and longevity. As such, an evaluation of the performance of modern cementless patellae is needed.
View Article and Find Full Text PDFCureus
December 2024
Trauma and Orthopaedics, Northampton General Hospital, Northampton, GBR.
Although mixing and matching components is a common, safe, and well-documented practice in hip revision surgery, our extensive search indicates that it has not been previously reported for shoulder arthroplasty. This case report presents the use of mixed implants in shoulder revision surgery to reduce morbidity and address flaws in the initial implant design. We describe a case of a patient with multiple epiphyseal dysplasia who was treated for osteoarthritis in his left shoulder with an anatomic shoulder replacement in 2014.
View Article and Find Full Text PDFJ Clin Orthop Trauma
January 2025
Loyola University Health System, Department of Orthopaedic Surgery and Rehabilitation, 2160 South First Avenue, 60153, Maywood, IL, USA.
Introduction: Mobile-bearing (MB) inserts, designed to minimize aseptic loosening and to reduce contact stresses leading to polyethylene wear, are an alternative to fixed-bearing (FB) inserts. Most studies have shown no significant difference between MB and FB constructs, and there is limited long-term data comparing the two constructs [1,2,3,4]. The purpose of this study was to report the outcomes of a randomized controlled trial comparing MB versus FB inserts on patients with minimum 20-year follow-up.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!