Purpose: In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA) based on the administered dose for adults in Japan. When the JSNM guidelines are used, the relationship between this recommended administered dose and image quality remains unclear. In this study, we clarified the relationship between administered dose and image quality for pediatric Tc-DMSA renal scan retrospectively, and verified the diagnosable image quality with the recommended administered dose of the JSNM guidelines.

Materials And Methods: Data from 7 pediatric patients who underwent Tc-DMSA dynamic renal scans according to the guidelines' recommended doses were collected. Scan frame rate was 1 frame/min, and scan time was up to 8 min. Eight images, which had different acquired time periods from 1 min to 8 min were prepared by adding each frame. Nine nuclear medicine specialists determined 8 images with different acquired times as diagnosable or undiagnosable. A region of interest (ROI) with 50% thresholds was placed on each kidney of every image. Coefficient of variation (CV) was calculated by dividing the standard deviation (σ) by the mean counts (µ) of each ROI (CV = σ/µ × 100). Tc-DMSA renal scans (total of 2821 cases) that were performed previously in collaboration with 6 hospitals were collected, and CVs of these images were calculated in all cases. These 2821 cases were separated into 5 groups for every 10 kg weight; i.e., (1) less than 10 kg, (2) 10-19.9 kg, (3) 20-29.9 kg, (4) 30-39.9 kg, and (5) above 40 kg. Regression line of each group was analyzed in relation to the CV and administered dose. The CV at the point of intersection with the recommended dose range from the guideline was determined for each group. This CV value was considered as the estimated CV of the image obtained when the recommended dose of the guideline was used. Thus, if the CV was equal to or less than the estimated CV value, then the diagnostic image quality was deemed satisfactory.

Results: Average CV of the lower limit of diagnosable images in 7 cases as determined by 9 nuclear medicine specialists was 19.9%. Estimated CV was 21.2-24.2% in the group weighing < 10 kg (group 1), 19.9-20.6% in the group weighing > 10 kg and < 20 kg (group 2), 19.6% in group weighing > 20 kg and < 30 kg (group 3), 19.4-19.5% in the group weighing > 30 kg and < 40 kg (group 4), and 19.8% in the group weighing > 40 kg (group 5). The estimated CVs from groups 1 and 2 with weight < 20 kg exceeded 19.9%.

Conclusions: Although Tc-DMSA renal scan can be carried out using the guidelines' recommended dose with conventional image acquisition time in patients weighing 20 kg or more, those < 20 kg need consideration for a longer image acquisition time to obtain diagnosable images.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12149-018-1320-6DOI Listing

Publication Analysis

Top Keywords

nuclear medicine
28
administered dose
24
image quality
20
dose image
12
tc-dmsa renal
12
japanese society
12
society nuclear
12
relation administered
8
dose
8
quality pediatric
8

Similar Publications

Development of a capsid protein-based ELISA for the detection of PCV2 antibodies in swine serum.

Pol J Vet Sci

December 2024

Key Laboratory of Animal Pathogen and Biosafety Education of the Ministry of Education, Zhengzhou 450000, China.

Porcine circovirus type 2 (PCV2) is the major causative agent of postweaning multisystemic wasting syndrome which leads to significant economic losses in the global swine industry. In China, there is a widespread dissemination of PCV2 infection in the pig population. Serological diagnosis of the disease is considered as an effective control measure.

View Article and Find Full Text PDF

Identification and Characterisation of Potential Targets for N6-methyladenosine (m6A) Modification during Intervertebral Disc Degeneration.

Front Biosci (Landmark Ed)

November 2024

Department of Orthopaedics, The Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University, 646000 Luzhou, Sichuan, China.

Background: The mechanism for RNA methylation during disc degeneration is unclear. The aim of this study was to identify N6-methyladenosine (m6A) markers and therapeutic targets for the prevention and treatment of intervertebral disc degeneration (IDD).

Methods: Methylated RNA immunoprecipitation sequencing (MeRIP-seq) and quantitative reverse transcription PCR (RT-qPCR) were employed to analyze m6A modifications of IDD-related gene expression.

View Article and Find Full Text PDF

Nanoscale surface topography is an effective approach in modulating cell-material interactions, significantly impacting cellular and nuclear morphologies, as well as their functionality. However, the adaptive changes in cellular metabolism induced by the mechanical and geometrical microenvironment of the nanotopography remain poorly understood. In this study, we investigated the metabolic activities in cells cultured on engineered nanopillar substrates by using a label-free multimodal optical imaging platform.

View Article and Find Full Text PDF

Background: Platinum chemotherapy (CT) remains the backbone of systemic therapy for patients with small-cell lung cancer (SCLC). The nucleotide excision repair (NER) pathway plays a central role in the repair of the DNA damage exerted by platinum agents. Alteration in this repair mechanism may affect patients' survival.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!