Objectives: The purpose of the present study was to compare the mean preparation times and frequency with which Reciproc and Reciproc Blue instruments reached the full working length in mandibular molars, with or without glide path preparation.

Materials And Methods: Previously untreated mesiobuccal and mesiolingual canals with completely formed apices were randomly divided into 6 groups ( = 50) depending on the usage of Reciproc (RC; VDW), Reciproc Blue (RC Blue; VDW), C-Pilot (CP; VDW), and R-Pilot (RP; VDW) files: RC, RC Blue, RC + C-Pilot (RC-CP), RC-Blue + C-Pilot (RC Blue-CP), RC+R-Pilot (RC-RP), and RC Blue + R-Pilot (RC Blue-RP). A glide path was prepared using the hand-operated C-Pilot or the machine-operated R-Pilot instruments, respectively. The χ test, analysis of variance, and the Tukey test were used for statistical comparisons.

Results: No statistically significant differences were observed in the distribution of the frequency of reaching the full working length in the RC (94%), RC Blue (88%), RC-CP (94%), RC Blue-CP (90%), RC-RP (96%), and RC Blue-RP (92%) groups ( > 0.05).

Conclusions: Preparation of a glide path did not have a significant effect on reaching the full working length using these systems.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6237731PMC
http://dx.doi.org/10.5395/rde.2018.43.e41DOI Listing

Publication Analysis

Top Keywords

full working
16
working length
16
glide path
16
reciproc blue
12
reciproc reciproc
8
blue instruments
8
reaching full
8
blue
7
reciproc
6
comparison ability
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!