Purpose: Since the publication of the AAPM TG128 report for the quality assurance (QA) of prostate brachytherapy ultrasound systems, no commercially available phantoms have been developed which satisfy all of the task group recommendations. Current solid phantoms require a separate user-implemented setup using a container with liquid medium to evaluate the alignment between the needle template and the electronic grid, a test of geometric accuracy with critical implications in dosimetric quality. Utilizing a 3D printer, we constructed a cost-effective, liquid-based phantom that provides a complete TG128 solution which improves the efficiency of brachytherapy ultrasound QA.

Methods: The TG128 report was used to guide the design process of the liquid-based phantom. The needle template and electronic grid alignment setup served as the foundation with specific components developed to integrate all remaining tests. Water was chosen as the liquid medium, with speed of sound adjusted to 1,540 m/s via salinity per the task group recommendations. The proof of concept was evaluated by comparing the time stamps labeled on QA images between the liquid-based phantom and a commercially available one for both a new and experienced user.

Results: A TG128 QA trial run demonstrated that all recommended tests can be completed with the single phantom setup. Evaluation of the time data revealed a total QA duration of 45 min (average of two trials) with the liquid-based phantom, compared to 70 and 90 min with the commercial phantom for a new and experienced user.

Conclusions: The liquid-based phantom is specifically designed to satisfy the recommendations of the TG128 report. The incorporation of 3D printing allows simple design modifications to adapt the phantom on-the-fly if needed. The resulting product improves the efficiency of brachytherapy ultrasound QA by eliminating the need for multiple phantom setups.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13307DOI Listing

Publication Analysis

Top Keywords

liquid-based phantom
20
brachytherapy ultrasound
16
tg128 report
12
phantom
10
single phantom
8
task group
8
group recommendations
8
liquid medium
8
needle template
8
template electronic
8

Similar Publications

The increasing utilization of cerebrospinal fluid (CSF) in early detection of Alzheimer's disease (AD) is attributed to the change of Amyloid- β proteins. Since, the brain is suspended in CSF, changes of Amyloid- β proteins in CSF reflect a pathophysiological variation of the brain due to AD. However, the correlation between Amyloid- β proteins and the dielectric properties (DPs) of CSF is still an open question.

View Article and Find Full Text PDF

Purpose: With external beam radiation therapy, uncertainties in treatment planning and delivery can result in an undesirable dose distribution delivered to the patient that can compromise the benefit of treatment. Techniques including geometric margins and probabilistic optimization have been used effectively to mitigate the effects of uncertainties. However, their broad application is inconsistent and can compromise the conclusions derived from cross-technique and cross-modality comparisons.

View Article and Find Full Text PDF

Purpose: Since the publication of the AAPM TG128 report for the quality assurance (QA) of prostate brachytherapy ultrasound systems, no commercially available phantoms have been developed which satisfy all of the task group recommendations. Current solid phantoms require a separate user-implemented setup using a container with liquid medium to evaluate the alignment between the needle template and the electronic grid, a test of geometric accuracy with critical implications in dosimetric quality. Utilizing a 3D printer, we constructed a cost-effective, liquid-based phantom that provides a complete TG128 solution which improves the efficiency of brachytherapy ultrasound QA.

View Article and Find Full Text PDF

Interstitial fiber-optic-based approaches used in both diagnostic and therapeutic applications rely on localized light-tissue interactions. We present an optical technique to identify spectrally and spatially specific exogenous chromophores in highly scattering turbid media. Point radiance spectroscopy is based on directional light collection at a single point with a side-firing fiber that can be rotated up to 360 deg.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!