Rationale And Objectives: Analysis of dynamic contrast-enhanced (DCE) magnetic resonance imaging is notable for the variability of calculated parameters. The purpose of this study was to evaluate the level of measurement variability and error/variability due to modeling in DCE magnetic resonance imaging parameters.

Materials And Methods: Two prostate DCE scans were performed on 11 treatment-naïve patients with suspected or confirmed prostate peripheral zone cancer within an interval of less than two weeks. Tumor-suspicious and normal-appearing regions of interest (ROI) in the prostate peripheral zone were segmented. Different Tofts-Kety based models and different arterial input functions, with and without bolus arrival time (BAT) correction, were used to extract pharmacokinetic parameters. The percent repeatability coefficient (%RC) of fitted model parameters K, v, and k was calculated. Paired t-tests comparing parameters in tumor-suspicious ROIs and in normal-appearing tissue evaluated each parameter's sensitivity to pathology.

Results: Although goodness-of-fit criteria favored the four-parameter extended Tofts-Kety model with the BAT correction included, the simplest two-parameter Tofts-Kety model overall yielded the best repeatability scores. The best %RC in the tumor-suspicious ROI was 63% for k, 28% for v and 83% for K . The best p values for discrimination between tissues were p <10 for k and K, and p = 0.11 for v. Addition of the BAT correction to the models did not improve repeatability.

Conclusion: The parameter k, using an arterial input functions directly measured from blood signals, was more repeatable than K. Both K and k values were highly discriminatory between healthy and diseased tissues in all cases. The parameter v had high repeatability but could not distinguish the two tissue types.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6526092PMC
http://dx.doi.org/10.1016/j.acra.2018.10.018DOI Listing

Publication Analysis

Top Keywords

arterial input
8
dynamic contrast-enhanced
8
dce magnetic
8
magnetic resonance
8
resonance imaging
8
prostate peripheral
8
peripheral zone
8
bat correction
8
tofts-kety model
8
selection fitting
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!