Objectives: The aim of this study was to investigate the hemodynamic and clinical performance of the Evolut PRO compared with its direct predecessor, the Evolut R.

Background: Recently, the newest commercially available generation of the self-expandable Medtronic CoreValve prosthesis, the CoreValve Evolut PRO, was introduced to the market. This prosthesis is based on the previous Evolut R model and specifically designed to mitigate paravalvular leakage. Because of the design changes, the Evolut PRO needs a larger sheath size (16-F vs. 14-F).

Methods: Patients receiving either the Evolut R (n = 148) or the Evolut PRO (n = 74) from September 2015 to January 2018 were compared in a 2:1 fashion after propensity score matching. Baseline characteristics, cardiovascular imaging, and pre- and periprocedural outcomes were prospectively collected and assessed.

Results: Both cohorts represent a high-risk, real-world collective with increased perioperative mortality risk (logistic European System for Cardiac Operative Risk Evaluation score, Evolut R vs. Evolut PRO: 24.7 ± 13.7% vs. 25.1 ± 12.5%; p = 0.881). Procedural success was 100%, and the mean transvalvular pressure gradient was substantially reduced (Evolut R vs. Evolut PRO: 7.9 ± 3.9 mm Hg vs. 7.5 ± 3.5 mm Hg; p = 0.348). Mild paravalvular leakage was observed in 16.2% of Evolut R patients and in 14.9% of Evolut PRO patients (p = 0.794). In the Evolut R group, moderate aortic regurgitation was documented in 2 patients (Evolut R vs. Evolut PRO: 1.4% vs. 0%; p = 1.000). No differences regarding clinical parameters, such as major bleeding events (Evolut R vs. Evolut PRO: 1.4% vs.1.3%; p = 0.868) and vascular complications were observed.

Conclusions: Both prostheses show excellent hemodynamic performance with a low incidence of paravalvular leakage and comparable clinical outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2018.07.023DOI Listing

Publication Analysis

Top Keywords

evolut pro
40
evolut
20
evolut evolut
16
paravalvular leakage
12
pro
10
pro 14%
8
p =
5
latest evolution
4
evolution medtronic corevalve
4
medtronic corevalve system
4

Similar Publications

A 90-year-old man received a diagnosis of ascending aortic pseudoaneurysms after transcatheter aortic valve implantation (TAVI) using an Evolut PRO valve (Medtronic). Plug closure of the pseudoaneurysms was successfully performed, and the symptoms improved after the procedure. However, on postoperative day 4, the patient experienced sudden massive hemoptysis and died.

View Article and Find Full Text PDF

Transcatheter Aortic Valve-in-Valve Implantation with Newer Generation Evolut Valve by Size of Failed Bioprosthesis.

Anatol J Cardiol

January 2025

Department of Cardiothoracic Surgery Research, Lankenau Institute for Medical Research, Wynnewood, Pennsylvania, USA ; Department of Cardiothoracic Surgery, Lankenau Heart Institute, Main Line Health Wynnewood, Pennsylvania, USA.

Background: To evaluate the clinical outcomes of valve-in-valve transcatheter aortic valve replacement (ViV TAVR) with newer-generation self-expanding Evolut valves according to the size of the failed surgical bioprosthesis.

Methods: This single-center retrospective study evaluated consecutive patients undergoing ViV TAVR with the Evolut Pro/Pro+/Fx between 2018 and 2022. These patients were compared based on the true internal diameter (ID) of the failed bioprosthesis, specifically ≤19 mm (small group) vs.

View Article and Find Full Text PDF

Background: The self-expanding, supra-annular Evolut valve is an established platform for Transcatheter Aortic Valve Implantation (TAVI). Evolut PRO introduced an outer sealing wrap to mitigate paravalvular leakage. We evaluated the 3-year clinical outcomes and valve performance of the Evolut PRO in standard clinical practice for severe aortic stenosis (AS) patients at intermediate or higher risk for surgery.

View Article and Find Full Text PDF

Background: No data compare newer-generation transcatheter heart valves (THVs) in terms of next-day discharge (NDD) following transfemoral (TF) transcatheter aortic valve implantation (TAVI).

Aims: We aimed to evaluate the safety of NDD in unselected patients who received ACURATE (neo/neo2), Evolut (PRO/PRO+/FX) and the SAPIEN (3/Ultra) THVs.

Methods: This multicentre registry included patients who underwent TF-TAVI without a preprocedural permanent pacemaker implantation (PPI) and were discharged the next day without a new PPI.

View Article and Find Full Text PDF

Background: Head-to-head comparisons of second and third generations of transcatheter heart valves (THVs) are mostly limited to 2-arm studies and to mid-term follow-up. The aim of this study was to simultaneously compare clinical outcomes of transcatheter aortic valve replacement (TAVR) with 4 different THVs at 5 years.

Methods: Patients undergoing transfemoral TAVR with 4 second-generation THV platforms and enrolled in the multicentre prospective OBSERVANT II study from December 2016 to September 2018 were compared according to the THV received.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!