Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability.
Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements. The Bland-Altman method was used as an additional tool to assess the level of agreement between the two algometers.
Results: The mean ± SD sensory thresholds measured with the EVF were 311 ± 116 g and 378 ± 178 g for the stifle and for the lumbosacral junction, respectively, whereas those measured with the SMALGO were 391 ±172 g and 476 ± 172 g. The inter-rater reliability was fair (ICC >0.4) for each pair of measurements except those taken at the level of the stifle with the SMALGO, for which the level of agreement between observers A and B was poor (ICC = 0.01). The inter-device reliability was good (ICC = 0.73; = 0.001). The repetition of the measurements affected reliability, as the thresholds obtained after the 45 min break were consistently lower than those measured during the first part of the trial ( = 0.02).
Conclusions And Relevance: The EVF and the SMALGO may be used interchangeably in cats, especially when the area to be tested is the lumbosacral joint. However, when the thresholds are measured at the stifle, the inter-observer reliability is better with the EVF than with the SMALGO. The reliability decreases when the measurements are repeated within a short time interval, suggesting a limited clinical applicability of quantitative sensory testing with both algometers in cats.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11132251 | PMC |
http://dx.doi.org/10.1177/1098612X18813426 | DOI Listing |
Sensors (Basel)
May 2023
School of Health and Life Sciences, Glasgow Caledonian University, Cowcaddens Road, Glasgow G4 0BA, UK.
A link between inappropriate physical behaviour patterns (low physical activity and high sedentary behaviour) and poor health outcomes has been observed. To provide evidence to quantify this link, it is important to have valid and reliable assessment tools. This study aimed to assess the validity and reliability of the activPAL4 monitor for distinguishing postures and measuring stepping activity of 6-12-year-old children.
View Article and Find Full Text PDFPOCUS J
November 2022
Department of General Practice, Faculty of Medicine, Sorbonne University Paris France.
The main objective of this study is the evaluation of the accuracy and reliability of a handheld point of care ultrasound device (POCUS-hd) for intrauterine pregnancy (IUP) detection compared to comprehensive reference transabdominal ultrasound (TU). The secondary objectives were to evaluate POCUS-hd for intrauterine pregnancy (IUP) detection compared to transabdominal and transvaginal ultrasound (TUTV), evaluate the inter-device agreement and inter-rater reliability of gestational age during early pregnancy. It is an observational transverse study with consecutive patient recruitment.
View Article and Find Full Text PDFTop Stroke Rehabil
March 2021
Laboratoire Analyse Et Restauration Du Mouvement (ARM), Hôpital Henri MONDOR, Université Paris-Est Créteil, Paris, France.
Background: One of the main challenges after stroke is gait recovery. To provide patients with an individualized rehabilitation program, it is helpful to have real-life objective evaluations at baseline and at regular follow-ups to adjust the program and verify potential improvements.
Objectives: To evaluate the accuracy and reliability of a fully stand-alone system of connected insoles (FeetMe® Monitor) against a widely used clinical walkway system (GAITRite®).
J Feline Med Surg
October 2019
Department of Clinical Sciences and Services, Royal Veterinary College, University of London, London, UK.
Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability.
Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements.
Eye Vis (Lond)
July 2018
1Department of Ophthalmology, Stanford University, 2370 Watson Court, suite 200, Palo Alto, CA 94303 USA.
Background: Optic nerve head measurements extracted from optical coherence tomography (OCT) show promise for monitoring clinical conditions with elevated optic nerve heads. The aim of this study is to compare reliability within and between raters and between image acquisition devices of optic nerve measurements derived from OCT scans in eyes with varying degrees of optic nerve elevation.
Methods: Wide angle line scans and narrow angle radial scans through optic nerve heads were obtained using three spectral domain(SD) OCT devices on 5 subjects (6 swollen optic nerves, 4 normal optic nerves).