Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability.

Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements. The Bland-Altman method was used as an additional tool to assess the level of agreement between the two algometers.

Results: The mean ± SD sensory thresholds measured with the EVF were 311 ± 116 g and 378 ± 178 g for the stifle and for the lumbosacral junction, respectively, whereas those measured with the SMALGO were 391 ±172 g and 476 ± 172 g. The inter-rater reliability was fair (ICC >0.4) for each pair of measurements except those taken at the level of the stifle with the SMALGO, for which the level of agreement between observers A and B was poor (ICC = 0.01). The inter-device reliability was good (ICC = 0.73; = 0.001). The repetition of the measurements affected reliability, as the thresholds obtained after the 45 min break were consistently lower than those measured during the first part of the trial ( = 0.02).

Conclusions And Relevance: The EVF and the SMALGO may be used interchangeably in cats, especially when the area to be tested is the lumbosacral joint. However, when the thresholds are measured at the stifle, the inter-observer reliability is better with the EVF than with the SMALGO. The reliability decreases when the measurements are repeated within a short time interval, suggesting a limited clinical applicability of quantitative sensory testing with both algometers in cats.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11132251PMC
http://dx.doi.org/10.1177/1098612X18813426DOI Listing

Publication Analysis

Top Keywords

inter-rater inter-device
12
inter-device reliability
12
electronic von
8
von frey
8
frey anaesthesiometer
8
healthy cats
8
sensory thresholds
8
stifle lumbosacral
8
lumbosacral joint
8
pair measurements
8

Similar Publications

A link between inappropriate physical behaviour patterns (low physical activity and high sedentary behaviour) and poor health outcomes has been observed. To provide evidence to quantify this link, it is important to have valid and reliable assessment tools. This study aimed to assess the validity and reliability of the activPAL4 monitor for distinguishing postures and measuring stepping activity of 6-12-year-old children.

View Article and Find Full Text PDF

The main objective of this study is the evaluation of the accuracy and reliability of a handheld point of care ultrasound device (POCUS-hd) for intrauterine pregnancy (IUP) detection compared to comprehensive reference transabdominal ultrasound (TU). The secondary objectives were to evaluate POCUS-hd for intrauterine pregnancy (IUP) detection compared to transabdominal and transvaginal ultrasound (TUTV), evaluate the inter-device agreement and inter-rater reliability of gestational age during early pregnancy. It is an observational transverse study with consecutive patient recruitment.

View Article and Find Full Text PDF

Background: One of the main challenges after stroke is gait recovery. To provide patients with an individualized rehabilitation program, it is helpful to have real-life objective evaluations at baseline and at regular follow-ups to adjust the program and verify potential improvements.

Objectives: To evaluate the accuracy and reliability of a fully stand-alone system of connected insoles (FeetMe® Monitor) against a widely used clinical walkway system (GAITRite®).

View Article and Find Full Text PDF

Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability.

Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements.

View Article and Find Full Text PDF

Background: Optic nerve head measurements extracted from optical coherence tomography (OCT) show promise for monitoring clinical conditions with elevated optic nerve heads. The aim of this study is to compare reliability within and between raters and between image acquisition devices of optic nerve measurements derived from OCT scans in eyes with varying degrees of optic nerve elevation.

Methods: Wide angle line scans and narrow angle radial scans through optic nerve heads were obtained using three spectral domain(SD) OCT devices on 5 subjects (6 swollen optic nerves, 4 normal optic nerves).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!