Objectives: Performance of a modified abdominopelvic CT protocol reconstructed using full iterative reconstruction (IR) was assessed for imaging patients presenting with acute abdominal symptoms.
Materials And Methods: Fifty-seven patients (17 male, 40 female; mean age of 56.5 ± 8 years) were prospectively studied. Low-dose (LD) and conventional-dose (CD) CTs were contemporaneously acquired between November 2015 and March 2016. The LD and CD protocols imparted radiation exposures approximating 10-20% and 80-90% those of routine abdominopelvic CT, respectively. The LD images were reconstructed with model-based iterative reconstruction (MBIR), and CD images with hybrid IR (40% adaptive statistical iterative reconstruction (ASIR)). Image quality was assessed quantitatively and qualitatively. Independent clinical interpretations were performed with a 6-week delay between reviews.
Results: A 74.7% mean radiation dose reduction was achieved: LD effective dose (ED) 2.38 ± 1.78 mSv (size-specific dose estimate (SSDE) 3.77 ± 1.97 mGy); CD ED 7.04 ± 4.89 mSv (SSDE 10.74 ± 5.5 mGy). LD-MBIR images had significantly lower objective and subjective image noise compared with CD-ASIR (p < 0.0001). Noise reduction for LD-MBIR studies was greater for patients with BMI < 25 kg/m than those with BMI ≥ 25 kg/m (5.36 ± 3.2 Hounsfield units (HU) vs. 4.05 ± 3.1 HU, p < 0.0001). CD-ASIR studies had significantly better contrast resolution, and diagnostic acceptability (p < 0.0001 for all). LD-MBIR studies had significantly lower streak artifact (p < 0.0001). There was no difference in sensitivity for primary findings between the low-dose and conventional protocols with the exception of one case of enteritis.
Conclusions: Low-dose abdominopelvic CT performed with MBIR is a feasible radiation dose reduction strategy for imaging patients presenting with acute abdominal pain.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10140-018-1658-z | DOI Listing |
Int J Comput Assist Radiol Surg
January 2025
Department of Medical Biophysics, University of Toronto, Toronto, Canada.
Purpose: During endovascular revascularization interventions for peripheral arterial disease, the standard modality of X-ray fluoroscopy (XRF) used for image guidance is limited in visualizing distal segments of infrapopliteal vessels. To enhance visualization of arteries, an image registration technique was developed to align pre-acquired computed tomography (CT) angiography images and to create fusion images highlighting arteries of interest.
Methods: X-ray image metadata capturing the position of the X-ray gantry initializes a multiscale iterative optimization process, which uses a local-variance masked normalized cross-correlation loss to rigidly align a digitally reconstructed radiograph (DRR) of the CT dataset with the target X-ray, using the edges of the fibula and tibia as the basis for alignment.
Phys Med Biol
January 2025
National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-ku, Chiba, 263-8555, JAPAN.
PET has become an important clinical modality but is limited to imaging positron emitters. Recently, PET imaging withZr, which has a half-life of 3 days, has attracted much attention in immuno-PET to visualize immune cells and cancer cells by targeting specific antibodies on the cell surface. However,Zr emits a single gamma ray at 909 keV four times more frequently than positrons, causing image quality degradation in conventional PET.
View Article and Find Full Text PDFPhys Med Biol
January 2025
Faculty of Mathematics and Natural Sciences , Hochschule Darmstadt, Schöfferstr., 3, Darmstadt, Hessen, 64295, GERMANY.
Magnetic Particle Imaging (MPI) is an emerging medical imaging modality which has gained increasing interest in recent years. Among the benefits of MPI are its high temporal resolution, and that the technique does not expose the specimen to any kind of ionizing radiation. It is based on the non-linear response of magnetic nanoparticles to an applied magnetic field.
View Article and Find Full Text PDFEur Radiol Exp
January 2025
Unit of Medical Physics, Pisa University Hospital "Azienda Ospedaliero-Universitaria Pisana", Pisa, Italy.
Background: Photon-counting detector (PCD) technology has the potential to reduce noise in computed tomography (CT). This study aimed to carry out a voxelwise noise characterization for a clinical PCD-CT scanner with a model-based iterative reconstruction algorithm (QIR).
Methods: Forty repeated axial acquisitions (tube voltage 120 kV, tube load 200 mAs, slice thickness 0.
Sci Rep
January 2025
Faculty of Physics, Department of Optics and Optometry and Vision Sciences, Universitat de València, Burjassot, Spain.
A lensless compact arrangement based on digital in-line holography under Gabor's regime is proposed as a novel contactless method to assess the profile of multifocal intraocular lenses (MIOLs) which are conformed by several diffractive rings. Diffractive MIOLs are a widely adopted ophthalmologic option for the correction of presbyopia in patients undergoing cataract surgery. The MIOL optical design might introduce non-negligible optical performance differences between lenses as well as the introduction of undesirable photic phenomena (such as halos and glare) affecting the vision of users.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!