This paper describes the effects of a novel functional electrical stimulation (FES) system which has been integrated in a powered exoskeleton to provide up to 10 channels of stimulation to users with paraplegia via surface electrodes. Experimental data collected from three users with spinal cord injury (SCI) indicate the system reduced the exoskeleton motor torques necessary to perform sit-to-stand transitions in the exoskeleton. All subjects exhibited reduced muscle spasticity immediately after walking in the exoskeleton with FES. Additionally, one subject with stretch-reflex spasms exhibited increased joint excursion and reduced exoskeleton motor torques required to achieve over-ground gait when FES was incorporated.

Download full-text PDF

Source
http://dx.doi.org/10.1109/EMBC.2018.8512810DOI Listing

Publication Analysis

Top Keywords

powered exoskeleton
8
reduced exoskeleton
8
exoskeleton motor
8
motor torques
8
exoskeleton
6
fes
4
fes coupled
4
coupled powered
4
exoskeleton cooperative
4
cooperative muscle
4

Similar Publications