Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: The distal femoral varization osteotomy (DFVO) by a lateral opening wedge osteotomy is an established intervention for patients suffering from lateral femoro-tibial osteoarthritis on a genu valgum deformity. In order to improve the accuracy of this correction, the use of a customized cutting guide (PSI) has been proposed as an alternative to conventional technique. The objective of our study was to compare the accuracy of post-operative alignment following DFVO in the coronal and sagittal plane using either a conventional abacus technique or PSI guide.
Method: Twenty-one patients that underwent lateral opening wedge osteotomy from a technique using PSI based on 3D CT-scans were matched 1:1 to 21 patients operated on using a conventional technique (pre-operative planning performed on standard radiographs). The accuracy of the correction was analyzed, comparing coronal and sagittal mechanical post-operative angles with pre-operative planning.
Results: With regard to alignment in the coronal plane (HKA correction), our study demonstrated a significant improvement in the accuracy of the correction obtained in the PSI group compared to the conventional group (0.43 ± 0.50 vs 3.95 ± 1.64 p < 0.001). In the sagittal plane (PDFA correction), we also found a significant improvement in correction accuracy in the PSI group (0.52 ± 0.60 vs 3.10 ± 1.83 p < 0.001). There was a significant decrease in operating time (delta 7.7 ± 3.07 (1.5-13.9) (p = 0.0.161) and fluoroscopic images taken (6.9 ± 0.54 (5.8-8) p < 0.001).
Conclusion: Our results suggest that the use of PSI in DFVO improves the accuracy of correction in both the coronal and sagittal planes compared to conventional techniques.