Indirect immunofluorescence and microneutralization methods for the detection of antibodies to poliovirus serotypes 1, 2, and 3 were compared. Of the 41 sera tested for poliovirus type 1 antibody, 40% were in complete agreement, 55% differed by one dilution, and 5% differed by two dilutions. For poliovirus type 2, 37 sera were tested; 56% completely agreed, and 44% differed by only one dilution. For poliovirus type 3, complete agreement occurred in 59% of 33 sera, while 41% differed by one dilution. No false-negative results were obtained. These findings suggest that indirect immunofluorescence for poliovirus types 1, 2, and 3 is as sensitive as the microneutralization method and could represent a less cumbersome alternative.