Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Athletes with an implantable cardioverter-defibrillator (ICD) may require unique optimal device-based tachycardia programming.
Objective: The purpose of this study was to assess the association of tachycardia programming characteristics of ICDs with occurrence of shocks, transient loss-of-consciousness, and death among athletes.
Methods: A subanalysis of a prospective, observational, international registry of 440 athletes with ICDs followed for a median of 44 months was performed. Programming characteristics were divided into groups for rate cutoff (very high, high, or low) and detection (long-detection interval [>nominal] or nominal). Endpoints included total, appropriate, and inappropriate shocks, transient loss-of-consciousness, and mortality.
Results: In this cohort, 62% were programmed with high-rate cutoff and 30% with long detection. No athlete died of an arrhythmia (related or unrelated) to ICD shocks. Three patients had sustained ventricular tachycardia below programmed detection rate, presenting as palpations and/or dizziness. ICD shocks were received by 98 athletes (64 appropriate, 32 inappropriate); 2 patients received both. Programming a high-rate cutoff was associated with decreased risk of total (P = .01) and inappropriate (P = .04) shocks overall and during competition or practice. Programming long-detection intervals was associated with fewer total shocks. Single- vs dual-chamber devices and the number of zones were unrelated to risk of shock. Transient loss-of-consciousness, associated with 27 appropriate shocks, was not related to programming characteristics.
Conclusion: High-rate cutoff and long-detection duration programming of ICDs in athletes at risk for sudden death can reduce total and inappropriate ICD shocks without affecting survival or the incidence of transient loss-of-consciousness.