Optimization of steady-state free precession MRI for lung ventilation imaging with F C F at 1.5T and 3T.

Magn Reson Med

POLARIS, Unit of Academic Radiology, Department of IICD, University of Sheffield, Sheffield, United Kingdom.

Published: February 2019

Purpose: To optimize F imaging pulse sequences for perfluoropropane (C F ) gas human lung ventilation MRI considering intrinsic in vivo relaxation parameters at both 1.5T and 3T.

Methods: Optimization of the imaging parameters for both 3D spoiled gradient (SPGR) and steady-state free precession (SSFP) F imaging sequences with inhaled 79% C F and 21% oxygen was performed. Phantom measurements were used to validate simulations of SNR. In vivo parameter mapping and sequence optimization and comparison was performed by imaging the lungs of a healthy adult volunteer. T and T mapping was performed in vivo to optimize sequence parameters for in vivo lung MRI. The performance of SSFP and SPGR was then evaluated in vivo at 1.5T and 3T.

Results: The in vivo T of C F was shown to be dependent upon lung inflation level (2.04 ms ± 36% for residual volume and 3.14 ms ± 28% for total lung capacity measured at 3T), with lower T observed near the susceptibility interfaces of the diaphragm and around pulmonary blood vessels. Simulation and phantom measurements indicate that a factor of ~2-3 higher SNR can be achieved with SSFP when compared with optimized SPGR. In vivo lung imaging showed a 1.7 factor of improvement in SNR achieved at 1.5T, while the theoretical improvement at 3T was not attained due to experimental SAR constraints, shorter in vivo T , and B inhomogeneity.

Conclusion: SSFP imaging provides increased SNR in lung ventilation imaging of C F demonstrated at 1.5T with optimized SSFP similar to the SNR that can be obtained at 3T with optimized SPGR.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6491987PMC
http://dx.doi.org/10.1002/mrm.27479DOI Listing

Publication Analysis

Top Keywords

lung ventilation
12
steady-state free
8
free precession
8
imaging
8
ventilation imaging
8
vivo
8
ssfp imaging
8
phantom measurements
8
vivo lung
8
snr achieved
8

Similar Publications

Purpose: Pulmonary MRI faces challenges due to low proton density, rapid transverse magnetization decay, and cardiac and respiratory motion. The fermat-looped orthogonally encoded trajectories (FLORET) sequence addresses these issues with high sampling efficiency, strong signal, and motion robustness, but has not yet been applied to phase-resolved functional lung (PREFUL) MRI-a contrast-free method for assessing pulmonary ventilation during free breathing. This study aims to develop a reconstruction pipeline for FLORET UTE, enhancing spatial resolution for three-dimensional (3D) PREFUL ventilation analysis.

View Article and Find Full Text PDF

Background: Despite the physiological advantages of positive end-expiratory pressure (PEEP), its optimal utilization during one-lung ventilation (OLV) remains uncertain. We aimed to investigate whether individualized PEEP titration by lung compliance is associated with a reduced risk of postoperative pulmonary complications during OLV.

Methods: We searched PubMed, Embase, and the Cochrane Central Register of Controlled Trials until April 1, 2024, to identify published randomized controlled trials that compared individualized PEEP titration by lung compliance with fixed PEEP during OLV.

View Article and Find Full Text PDF

Background: Foreign body inhalation is rare in older children, often leading to underdiagnosis and delayed treatment. Most cases involve a single foreign body, but instances of multiple foreign bodies are exceedingly uncommon. This report presents a case of an elder child who inhaled two pen caps, emphasizing the need for clinical vigilance and thorough medical history collection.

View Article and Find Full Text PDF

Unlabelled: To evaluate the accuracy of the lung ultrasound score (LUS) in predicting ventilatory weaning failure during neonatal hospitalization in the NICU and to identify factors associated with weaning failure, including corrected gestational age (CGA). This prospective, longitudinal, pragmatic and observational cohort study included neonates on mechanical ventilation for at least 48 h. The primary outcome was the accuracy of lung ultrasound in predicting 3-day weaning failure, with the ROC curve used to determine the best LUS cutoff (sensitivity and specificity).

View Article and Find Full Text PDF

Negative Pressure Ventilation Ex-Situ Lung Perfusion Preserves Porcine and Human Lungs for 36-Hours.

Clin Transplant

January 2025

Division of Cardiac Surgery, Department of Surgery, Faculty of Medicine, University of Alberta, Edmonton, Canada.

Introduction: Preclinically, 24-hour continuous Ex-Situ Lung Perfusion (ESLP) is the longest duration achieved in large animal models and rejected human lungs. Here, we present our 36-hour Negative Pressure Ventilation (NPV)-ESLP protocol applied to porcine and rejected human lungs.

Methods: Five sets of donor domestic pig lungs (45-55 kg) underwent 36-hour NPV-ESLP.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!