Instruments to reduce the risk of tibial fracture following cementless unicompartmental knee replacement.

Knee

Nuffield Department of Orthopaedics, Rheumatology and Musculoskeletal Sciences (NDORMS), University of Oxford, Oxford, UK.

Published: December 2018

Background: Tibial plateau fracture is an important complication of cementless Oxford unicompartmental knee replacement. The press fit between the keel and the bone's keel slot is responsible for primary fixation and also contributes to fracture risk. This study investigates whether the fracture risk could be reduced without compromising primary fixation, by using different instruments to widen the keel slot.

Methods: Keel slots were made in polyurethane blocks (n = 60) using the standard keel cut saw blade or a new blade that was 0.2 mm wider, with adjuvant use of the cemented pick or a prototype rasp. A tibial component was pushed into and pulled out of the slots using a Dartec materials testing machine. It was assumed that the 'push-in' force was related to the risk of fracture and 'pull-out' force was related to fixation. Reproducibility studies with 10 different tibial components were undertaken.

Results: The new blade required significantly lower push-in forces than the standard blade (789 N SD 130, 1411 N SD 180; P < 0.001), but the pull-out forces were not different (240 N SD 47, 230 N SD 56; P > 0.999). With the standard blade the pick decreased the push-in (818 N SD 318; P < 0.001) and pull-out (128 N SD 58; P < 0.001) forces, but the rasp had no effect. With the new blade the pick had no effect, but the rasp increased the push-in force (1390 N SD 202; P < 0.001).

Conclusions: This study suggests the fracture risk will be reduced with the new blades, with no compromise in fixation. If the new blades are used routine use of the cemented pick appears to be of no benefit.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.knee.2018.09.001DOI Listing

Publication Analysis

Top Keywords

unicompartmental knee
8
knee replacement
8
primary fixation
8
fracture risk
8
standard blade
8
fracture
5
keel
5
blade
5
instruments reduce
4
risk
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!