Effects of knee simulator control method and radiation dose on UHMWPE wear rate, and relationship between wear rate and clinical revision rate in National Joint Registry.

J Mech Behav Biomed Mater

JFE Techno-Research Corporation, 1-1 Minamiwatarida-cho, Kawasaki-ku, Kawasaki city, Kanagawa 210-0855, Japan.

Published: February 2019

The wear rate of five types of cruciate-retaining artificial knee joint ultrahigh-molecular-weight polyethylene (UHMWPE) inserts was examined using two custom-made knee joint simulators satisfying ISO 14243-1 (load control) and ISO 14243-3 (displacement control). The wear rate of knee joints composed of a UHMWPE insert and a Co-Cr-Mo alloy or oxidized zirconium femoral component linearly increased with increasing number of wear cycles, and the volumetric wear rate per million cycles was approximately 6-18 mm/Mc. The wear rate was the lowest in the highly crosslinked knee joint irradiated at 90 kGy (Scorpio) among the five UHMWPE inserts. The extent of oxidation in UHMWPE after a knee simulator test of up to 5 × 10 cycles was small. The load-controlled wear rates measured in this work were close to the displacement-controlled wear rates reported in the literature. The effect of the control method on the wear rate was small for Nexgen and Scorpio knee joints. However, it was larger for the PFC Sigma knee joint having a high curvature of the surface. The wear rate of various knee joints made of highly crosslinked UHMWPE (XLPE) markedly decreased when they were subjected to a radiation dose of 40 kGy or more. The 10-year cumulative percentage revision rate since the primary operation slightly decreased with decreasing volumetric knee simulator wear rate for conventional UHMWPE (CPE) and XLPE knee joint inserts. The XLPE knee joint was shown to exhibit reduced in vivo wear and lower rates of revision for total knee replacement. On the other hand, Nexgen and PFC Sigma (both CPE) knee joints showed the lowest revision rate in the AOA and NJR national joint replacement registries. The volumetric wear rates of 3 mm/Mc for XLPE and 15 mm/Mc for CPE knee joint inserts are recommended as a goal for the development of new knee joints.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2018.09.034DOI Listing

Publication Analysis

Top Keywords

wear rate
36
knee joint
28
knee joints
20
knee
15
wear
14
knee simulator
12
rate
12
revision rate
12
wear rates
12
joint
9

Similar Publications

Physical activity (PA) reduces the risk of negative mental and physical health outcomes in older adults. Traditionally, PA intensity is classified using METs, with 1 MET equal to 3.5 mL O·min·kg.

View Article and Find Full Text PDF

Seroprevalence of SARS-COV-2 Antibodies Associated with the Use of Personal Protective Equipment in Healthcare Workers.

Mater Sociomed

January 2024

Department of Occupational Medicine, Faculty of Medicine, University of Tuzla, Tuzla, Bosnia and Herzegovina.

Background: Hypothetically, PPE played an estimated influential role in preventing SARS-CoV-2 transmission among HCWs before immune-prophylaxis by vaccination in BH. Objective: This research aims to determine the relationship between PPE use and the serological response to SARS-COV-2 among HCWs.

Methods: The sample contained 127 COVID-19 outpatients with an average age of 43.

View Article and Find Full Text PDF

Background: Postoperative nausea and vomiting (PONV) is the most common complication following general anesthesia. Currently, pharmaceutical therapy is the primary method of treatment, but it has reached a plateau, and it is accompanied by inherent adverse reactions and high costs. Stimulation of the wrist acupuncture point PC6 is recommended as an effective means of preventing PONV.

View Article and Find Full Text PDF

[Cataracts extraction combined with multifocal intraocular lens optic implantation in Berger space in school-age children with cataracts].

Zhonghua Yan Ke Za Zhi

January 2025

Eye Institute of Shandong First Medical University, Qingdao Eye Hospital of Shandong First Medical University, State Key Laboratory Cultivation Base, Shandong Key Laboratory of Eye Diseases, School of Ophthalmology, Shandong First Medical University, Qingdao266071, China.

To evaluate the safety and efficacy of cataract extraction combined with multifocal intraocular lens (IOL) optic implantation in Berger space in school-age children with cataracts. It was a prospective study. The clinical data of school-age children with cataract who underwent cataract extraction combined with multifocal IOL implantation at Qingdao Eye Hospital of Shandong First Medical University from January 2019 to June 2023 were collected.

View Article and Find Full Text PDF

Associations between choroidal thickness and rate of axial elongation in orthokeratology lens users.

Photodiagnosis Photodyn Ther

December 2024

Joint Shantou International Eye Center of Shantou University and The Chinese, University of Hong Kong, Shantou, Guangdong 515041, China. Electronic address:

Objective: To investigate the impact of using orthokeratology lenses (OK lenses) for one year on the axial length (AL) and choroidal thickness (ChT) in the 6 mm concentric central retinal region.

Methods: 36 myopic children (36 eyes) aged 8 to 16 years were enrolled. For the duration of one year, the OK group (18 subjects) utilized OK lenses, while the spectacles group (18 subjects) utilized single-vision spectacles as a control.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!