Purpose: Investigate the effective performance and safety of a new hypersonic vitrector technology.

Methods: Postapproval, prospective, single-arm, noncomparative, open-label study at one clinical site in India.

Indications: macular hole (9/20), vitreous hemorrhage (7/20), vitreomacular traction (3/20), and vitreomacular traction with pseudomacular hole (1/20). Safety endpoints included intraoperative and postoperative adverse events. Effective performance endpoints were surgeon-rated effectiveness, range of surgical time, and device settings. Other performance measures were preoperative and postoperative best-corrected visual acuity, slit-lamp and indirect ophthalmoscopy, applanation tonometry, color fundus photography, fundus fluorescein angiography, and spectral domain optical coherence tomography.

Results: Core vitreous removal (20/20 subjects), peripheral vitreous removal (18/20), and posterior vitreous detachment induction (13/15) surgeries were successfully completed. Total surgical time was 22.5 minutes to 106 minutes. Serious adverse events through 3 months were 2 device-associated retinal tears and detachment (one intraoperative) and one unrelated postoperative enlargement of macular hole with subretinal fluid.

Conclusion: This first-in-human study suggests that this new hypersonic vitrector technology is a promising alternative to commercially available guillotine vitrectors. The hypersonic vitrector was effective in core vitreous removal in all cases. Larger-scale studies are required to expand on our initial findings for induction of a posterior vitreous detachment or peripheral vitrectomy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6924930PMC
http://dx.doi.org/10.1097/IAE.0000000000002365DOI Listing

Publication Analysis

Top Keywords

hypersonic vitrector
12
vitreous removal
12
effective performance
8
macular hole
8
vitreomacular traction
8
adverse events
8
surgical time
8
core vitreous
8
posterior vitreous
8
vitreous detachment
8

Similar Publications

In our report, we present the hypersonic vitrectomy (Vitesse, Bausch and Lomb) being employed for anterior vitreous liquefaction and removal in posterior capsular rupture. The capsular tear with nucleus drop during conventional phacoemulsification was managed by vitrectomy using the hypersonic vitrector after posterior-assisted levitation followed by intraocular lens (IOL) implantation. The minimal cortical and epinuclear lens particles in the anterior chamber and vitreous were also liquefied with a stoke length of 30 to 40 μm and aspirated via the Vitesse vitrectomy system.

View Article and Find Full Text PDF

Purpose: Investigate the effective performance and safety of a new hypersonic vitrector technology.

Methods: Postapproval, prospective, single-arm, noncomparative, open-label study at one clinical site in India.

Indications: macular hole (9/20), vitreous hemorrhage (7/20), vitreomacular traction (3/20), and vitreomacular traction with pseudomacular hole (1/20).

View Article and Find Full Text PDF

Introduction: Qualitatively assess the possible delayed structural, macroscopic and microscopic changes in the neuro-retina, retinal vasculature, retinal pigment epithelium (RPE) and optic nerve head (ONH) after pars plana vitrectomy (PPV) surgery using a new hypersonic vitrector (HV).

Materials And Methods: Eight live porcine eyes underwent PPV using either the HV or a conventional pneumatic guillotine vitrector (GV). The un-operated fellow eye from each pig was used as an external control.

View Article and Find Full Text PDF

Performance analysis of a new hypersonic vitrector system.

PLoS One

September 2017

Manchester Royal Eye Hospital, Central Manchester University Hospitals NHS Foundation Trust, Manchester Academic Health Science Centre, Manchester, United Kingdom.

Purpose: To evaluate porcine vitreous flow and water flow rates in a new prototype hypersonic vitrectomy system compared to currently available pneumatic guillotine vitrectors (GVs) systems.

Methods: Two vitrectors were tested, a prototype, ultrasound-powered, hypersonic vitrector (HV) and a GV. Porcine vitreous was obtained within 12 to 24 h of sacrifice and kept at 4°C.

View Article and Find Full Text PDF

Purpose: Preliminary assessment of a new prototype ultrasound-based hypersonic vitrector (HV) by qualitatively examining the histopathological changes in the retina and vitreous body after pars plana vitrectomy (PPV) and its ability to fragment vitreous collagen.

Methods: Fourteen porcine cadaveric eyes, 20 eyes in live swine and six human cadaveric eyes underwent PPV using the HV or a pneumatic guillotine vitrector (GV). An additional 4 porcine crystalline lenses were touched with either the HV or GV for 1 minute.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!