Background: Quantitative diffusion-weighted imaging (DWI) MRI is a promising technique for cancer characterization and treatment monitoring. Knowledge of the reproducibility of DWI metrics in breast tumors is necessary to apply DWI as a clinical biomarker.

Purpose: To evaluate the repeatability and reproducibility of breast tumor apparent diffusion coefficient (ADC) in a multi-institution clinical trial setting, using standardized DWI protocols and quality assurance (QA) procedures.

Study Type: Prospective.

Subjects: In all, 89 women from nine institutions undergoing neoadjuvant chemotherapy for invasive breast cancer.

Field Strength/sequence: DWI was acquired before and after patient repositioning using a four b-value, single-shot echo-planar sequence at 1.5T or 3.0T.

Assessment: A QA procedure by trained operators assessed artifacts, fat suppression, and signal-to-noise ratio, and determine study analyzability. Mean tumor ADC was measured via manual segmentation of the multislice tumor region referencing DWI and contrast-enhanced images. Twenty cases were evaluated multiple times to assess intra- and interoperator variability. Segmentation similarity was assessed via the Sørenson-Dice similarity coefficient.

Statistical Tests: Repeatability and reproducibility were evaluated using within-subject coefficient of variation (wCV), intraclass correlation coefficient (ICC), agreement index (AI), and repeatability coefficient (RC). Correlations were measured by Pearson's correlation coefficients.

Results: In all, 71 cases (80%) passed QA evaluation: 44 at 1.5T, 27 at 3.0T; 60 pretreatment, 11 after 3 weeks of taxane-based treatment. ADC repeatability was excellent: wCV = 4.8% (95% confidence interval [CI] 4.0, 5.7%), ICC = 0.97 (95% CI 0.95, 0.98), AI = 0.83 (95% CI 0.76, 0.87), and RC = 0.16 * 10 mm /sec (95% CI 0.13, 0.19). The results were similar across field strengths and timepoint subgroups. Reproducibility was excellent: interreader ICC = 0.92 (95% CI 0.80, 0.97) and intrareader ICC = 0.91 (95% CI 0.78, 0.96).

Data Conclusion: Breast tumor ADC can be measured with excellent repeatability and reproducibility in a multi-institution setting using a standardized protocol and QA procedure. Improvements to DWI image quality could reduce loss of data in clinical trials.

Level Of Evidence: 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:1617-1628.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6524146PMC
http://dx.doi.org/10.1002/jmri.26539DOI Listing

Publication Analysis

Top Keywords

repeatability reproducibility
16
dwi
8
breast tumor
8
setting standardized
8
tumor adc
8
adc measured
8
reproducibility
6
95%
6
adc
5
breast
5

Similar Publications

A New Approach to Non-Invasive Microcirculation Monitoring: Quantifying Capillary Refill Time Using Oximetric Pulse Waves.

Sensors (Basel)

January 2025

Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, 168 Litang Road, Changping District, Beijing 102218, China.

(1) Background: To develop a novel capillary refill time measurement system and evaluate its reliability and reproducibility. (2) Methods: Firstly, the utilization of electromagnetic pressure technology facilitates the automatic compression and instantaneous release of the finger. Secondly, the employment of pressure sensing technology and photoelectric volumetric pulse wave analysis technology enables the dynamic monitoring of blood flow in distal tissues.

View Article and Find Full Text PDF

Precision of Fungal Resistance Test Method for Cereal Husk-Reinforced Composite Construction Profiles Considering Mycelium Removal Techniques.

Materials (Basel)

January 2025

Construction Materials Engineering Department, Instytut Techniki Budowlanej, 00-611 Warsaw, Poland.

Many test methods used in the assessment of construction products are characterised by poor precision, which is reflected in the discrepancies of results obtained by different laboratories. The test procedure for fungal resistance of natural particle-reinforced composite construction profiles has not been fully specified, which may lead to such discrepancies and erroneous utility assessment. In this study, the precision of the method and the influence of the mycelium removal on the fungal resistance were assessed based on the flexural strength and modulus of elasticity test results obtained for millet- and oat husk-reinforced PVC composites exposed to .

View Article and Find Full Text PDF

One-week test-retest stability of heart rate variability during rest and deep breathing.

Physiol Meas

January 2025

Department for Psychosomatic Medicine and Psychotherapy, Universitatsklinikum Jena, Philosophenweg 3, Jena, Thüringen, 07743, GERMANY.

Heart rate variability (HRV) is an important indicator of cardiac autonomic function. Given its clinical significance, reliable HRV assessment is crucial. Here, we assessed test-retest stability, as a key aspect of reliability, quantifying the consistency of a measure when repeated under the same conditions.

View Article and Find Full Text PDF

Background: Virtual patients (VPs) are computer screen-based simulations of patient-clinician encounters. VP use is limited by cost and low scalability.

Objective: Show proof-of-concept that VPs powered by large language models (LLMs) generate authentic dialogs, accurate representations of patient preferences, and personalized feedback on clinical performance; and explore LLMs for rating dialog and feedback quality.

View Article and Find Full Text PDF

Purpose: Evaluating sacral fractures is crucial in fragility fractures of the pelvis. Dual-energy CT (DECT) is considered useful for diagnosing unclear fractures on single-energy CT (SECT). This study aims to investigate the effectiveness of DECT in diagnosing sacral fractures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!