Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To evaluate the effect of intentional undersizing of prosthetic radial head implant diameters on joint contact pressures.
Methods: Eight fresh-frozen cadaveric elbows were aligned in neutral extension and loaded with 100 N using a custom testing apparatus. Radiocapitellar contact pressures were recorded using a Tekscan thin-film pressure sensor. Prosthetic radial head replacement was performed with 2 prostheses: the Anatomic Radial Head and the Evolve Proline Radial Head prostheses. Each design was sized according to the manufacturer's recommendations and then again using 2-mm smaller radial heads.
Results: Average and peak pressures were significantly higher with the Evolve than the Anatomic prostheses (P < 0.03 and 0.02, respectively). Peak pressures decreased from 4.2 ± 0.5 MPa to 2.9 ± 0.3 MPa for the Anatomic Radial Heads and from 5.6 ± 0.5 MPa to 3.9 ± 0.6 MPa when the Evolve Radial Heads were undersized by 2 mm. The mean pressures of the Anatomic Radial Heads (1.4 ± 0.1 MPa) did not change significantly with undersizing (1.3 ± 0.1 MPa, P = 0.12), whereas the mean pressures of the Evolve Radial Heads (1.6 ± 0.1 MPa) were significantly reduced with undersizing (1.4 ± 0.1 MPa, P < 0.02).
Conclusion: Both mean and peak pressures were initially high for the Evolve Radial Head sized based on the short axis diameter and were improved with further undersizing by 2 mm. Peak, but not mean, contact pressures were improved by undersizing the Anatomic prosthesis based on the long axis diameter.
Clinical Relevance: These findings support the clinical recommendation of some surgeons to undersize the Evolve prosthesis by 2-mm smaller diameter than the current manufacturer's suggestion and give reason to consider doing the same for the Anatomic prosthesis.