Purpose: This study developed and validated a Motion Artifact Quantification algorithm to automatically quantify the severity of motion artifacts on coronary computed tomography angiography (CCTA) images. The algorithm was then used to develop a Motion IQ Decision method to automatically identify whether a CCTA dataset is of sufficient diagnostic image quality or requires further correction.
Method: The developed Motion Artifact Quantification algorithm includes steps to identify the right coronary artery (RCA) regions of interest (ROIs), segment vessel and shading artifacts, and to calculate the motion artifact score (MAS) metric. The segmentation algorithms were verified against ground-truth manual segmentations. The segmentation algorithms were also verified by comparing and analyzing the MAS calculated from ground-truth segmentations and the algorithm-generated segmentations. The Motion IQ Decision algorithm first identifies slices with unsatisfactory image quality using a MAS threshold. The algorithm then uses an artifact-length threshold to determine whether the degraded vessel segment is large enough to cause the dataset to be nondiagnostic. An observer study on 30 clinical CCTA datasets was performed to obtain the ground-truth decisions of whether the datasets were of sufficient image quality. A five-fold cross-validation was used to identify the thresholds and to evaluate the Motion IQ Decision algorithm.
Results: The automated segmentation algorithms in the Motion Artifact Quantification algorithm resulted in Dice coefficients of 0.84 for the segmented vessel regions and 0.75 for the segmented shading artifact regions. The MAS calculated using the automated algorithm was within 10% of the values obtained using ground-truth segmentations. The MAS threshold and artifact-length thresholds were determined by the ROC analysis to be 0.6 and 6.25 mm by all folds. The Motion IQ Decision algorithm demonstrated 100% sensitivity, 66.7% ± 27.9% specificity, and a total accuracy of 86.7% ± 12.5% for identifying datasets in which the RCA required correction. The Motion IQ Decision algorithm demonstrated 91.3% sensitivity, 71.4% specificity, and a total accuracy of 86.7% for identifying CCTA datasets that need correction for any of the three main vessels.
Conclusion: The Motion Artifact Quantification algorithm calculated accurate (<10% error) motion artifact scores using the automated segmentation methods. The developed algorithms demonstrated high sensitivity (91.3%) and specificity (71.4%) in identifying datasets of insufficient image quality. The developed algorithms for automatically quantifying motion artifact severity may be useful for comparing acquisition techniques, improving best-phase selection algorithms, and evaluating motion compensation techniques.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.13243 | DOI Listing |
Photoacoustic microscopy has demonstrated outstanding performance in high-resolution functional imaging. However, in the process of photoacoustic imaging, the photoacoustic signals will be polluted by inevitable background noise. Besides, the image quality is compromised due to the biosafety limitation of the laser.
View Article and Find Full Text PDFJMIR Form Res
January 2025
Department of Computer Science, University of California, Irvine, Irvine, CA, United States.
Background: Acute pain management is critical in postoperative care, especially in vulnerable patient populations that may be unable to self-report pain levels effectively. Current methods of pain assessment often rely on subjective patient reports or behavioral pain observation tools, which can lead to inconsistencies in pain management. Multimodal pain assessment, integrating physiological and behavioral data, presents an opportunity to create more objective and accurate pain measurement systems.
View Article and Find Full Text PDFIn image-guided radiotherapy (IGRT), four-dimensional cone-beam computed tomography (4D-CBCT) is critical for assessing tumor motion during a patients breathing cycle prior to beam delivery. However, generating 4D-CBCT images with sufficient quality requires significantly more projection images than a standard 3D-CBCT scan, leading to extended scanning times and increased imaging dose to the patient. To address these limitations, there is a strong demand for methods capable of reconstructing high-quality 4D-CBCT images from a 1-minute 3D-CBCT acquisition.
View Article and Find Full Text PDFJ Med Imaging (Bellingham)
January 2025
Case Western Reserve University, Department of Biomedical Engineering, Cleveland, Ohio, United States.
Purpose: We investigated the feasibility and advantages of using non-contrast CT calcium score (CTCS) images to assess pericoronary adipose tissue (PCAT) and its association with major adverse cardiovascular events (MACE). PCAT features from coronary computed tomography angiography (CCTA) have been shown to be associated with cardiovascular risk but are potentially confounded by iodine. If PCAT in CTCS images can be similarly analyzed, it would avoid this issue and enable its inclusion in formal risk assessment from readily available, low-cost CTCS images.
View Article and Find Full Text PDFMagn Reson Imaging
January 2025
Department of Medical Imaging, Pingyin people's Hospital, Jinan 250400, China.
Magnetic Resonance Imaging is a cornerstone of medical diagnostics, providing high-quality soft tissue contrast through non-invasive methods. However, MRI technology faces critical limitations in imaging speed and resolution. Prolonged scan times not only increase patient discomfort but also contribute to motion artifacts, further compromising image quality.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!