Background: The objective of this study was to determine the impact of the type and orientation of peripheral screw placement in an eccentrically positioned locking plate on the structural integrity of the plate-diaphyseal bone interface. We hypothesized that central placement of the screw at the end of the plate in this setting is more important than screw type (locking versus nonlocking) to limiting the risk of subsequent fracture.

Methods: Twenty osteoporotic fourth-generation composite left humeri were divided into 4 groups and plated with stainless-steel 6-hole locking plates and 4.5-mm screws. Group 1 (control group) consisted of a centrally positioned plate with a centrally placed non-locking end screw at the sixth, most-proximal hole. Group 2 consisted of an eccentrically positioned plate with a non-locking proximal end screw placed through the center of the bone. Group 3 consisted of an eccentrically positioned plate with a locking proximal end screw placed perpendicular to the plate and eccentrically across the cortex. Group 4 consisted of an eccentrically positioned plate with a non-locking proximal end screw placed perpendicular to the plate and eccentrically across the cortex. Each group was tested with a single load to failure in torsion at a rate of 1°/second.

Results: The control group (Group 1) failed at significantly higher peak torque values (51.62 ± 7.35 Nm) than Group 2 (38.98 ± 6.78 Nm; p = 0.006), Group 3 (34.75 ± 1.81 Nm; p < 0.001), and Group 4 (31.55 ± 1.23 Nm; p < 0.001). Failure energy absorbed in Group 1 (2,591.49 ± 819.63 Nm/degree) was significantly higher than Group 3 (1,430.51 ± 449.99 Nm/degree; p = 0.04) and Group 4 (952.49 ± 123.52 Nm/degree; p = 0.004), but not significantly higher than Group 2 (1,847.73 ± 827.35 Nm/degree; p = 0.27).

Conclusions: Eccentrically placed plating of humeral shaft fractures significantly increases the risk of peri-implant fracture compared with a centrally placed plate. Directing the proximal-end screw centrally in an eccentrically placed plate may help to mitigate this risk at the proximal end.

Clinical Relevance: When possible, care should be taken to place the plate centrally on the bone to avoid increased risk of peri-implant fracture at the proximal plate-bone interface.

Download full-text PDF

Source
http://dx.doi.org/10.2106/JBJS.18.00024DOI Listing

Publication Analysis

Top Keywords

eccentrically positioned
16
group consisted
16
positioned plate
16
group
15
plate
12
risk peri-implant
12
peri-implant fracture
12
consisted eccentrically
12
proximal screw
12
screw
9

Similar Publications

Background: The purpose of this study was to assess impingement-free internal rotation (IR) in a virtual reverse shoulder arthroplasty simulation using a Statistical Shape Model based on scapula size.

Methods: A database of over 10,000 scapulae utilized for preoperative planning for shoulder arthroplasty was analyzed with a Statistical Shape Model to obtain 5 scapula sizes including the mean and 2 standard deviations. For each scapula model, one glenosphere size (33-42 mm) was selected as the best fit based on consensus among 3 shoulder surgeons.

View Article and Find Full Text PDF

Robot calibration and modelling measurements are commonly performed using a laser tracker. To capture three-dimensional positions, a SMR is attached to the robot. While some researchers employ adhesive bonds for this purpose, such methods often result in inaccurate, unstable and non-repeatable SMR positioning, adversely affecting measurement precision and the traceability of research outcomes.

View Article and Find Full Text PDF

Preservation and early evolution of scalidophoran ventral nerve cord.

Sci Adv

January 2025

State Key Laboratory of Continental Dynamics, Shaanxi Key Laboratory of Early Life & Environments and Department of Geology, Northwest University, Xi'an, China.

Ecdysozoan worms (Nematoida + Scalidophora) are typified by disparate grades of neural organization reflecting a complex evolutionary history. The fossil record offers a unique opportunity to reconstruct the early character evolution of the nervous system via the exceptional preservation of extinct representatives. We focus on their nervous system as it appears in early and mid-Cambrian fossils.

View Article and Find Full Text PDF

Comparative Analysis of Physiological Vergence Angle Calculations from Objective Measurements of Gaze Position.

Sensors (Basel)

December 2024

Department of Optometry and Vision Science, Faculty of Science and Technology, University of Latvia, Jelgavas Street 1, LV-1004 Riga, Latvia.

Eccentric photorefractometry is widely used to measure eye refraction, accommodation, gaze position, and pupil size. While the individual calibration of refraction and accommodation data has been extensively studied, gaze measurements have received less attention. PowerRef 3 does not incorporate individual calibration for gaze measurements, resulting in a divergent offset between the measured and expected gaze positions.

View Article and Find Full Text PDF

Spatial updating, the ability to track the egocentric position of surrounding objects during self-motion, is fundamental to navigating around the world. However, people make systematic errors when updating the position of objects after linear self-motion. To determine the source of these errors, we measured errors in remembered target position with or without passive lateral translations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!