Background: One of the most significant Intensity Modulated Radiation Therapy treatment benefits is a high target to normal tissue dose ratio. To improve this advantage, an additional accessory such as a compensator is used to deliver doses. Compensator-based IMRT treatment is usually operated with an energy higher than 10 MV. Photoneutrons, which have high linear energy transfer and radiobiological effectiveness, are produced by colliding high-energy photon beams with linear accelerator structures, then they deliver the unwanted doses to patients and staff. Therefore, the neutron energy spectra should be determined in order to calculate and reduce the photoneutron risk.

Objective: We have conducted a comprehensive and precise study on the influence of brass compensator thickness and field size on neutron contamination spectrum in an Elekta SL 75/25 medical linear accelerator with and without the flattening filter by Monte Carlo method.

Material And Methods: MCNPX MC Code version 2.6.0 was utilized to simulate the detailed geometry of Elekta SL 75/25 head components based on Linac's manual. This code includes an important feature to simulate the photo-neutron interactions. Photoneutrons spectrum was calculated after the Linac output benchmarking based on tuning the primary electron beam.

Results And Conclusion: Based on the Friedman and Wilcoxon nonparametric tests results (P<0.05), photoneutron fluence directly depends on the field size and compensator thickness. Moreover, the unflattened beam provides lower photoneutron fluence than the flattened beam. Photoneutrons fluence is not negligible in compensator-based IMRT treatment. However, in order to optimize treatment plans, this additional and unwanted dose must be accounted for patientss.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6169126PMC

Publication Analysis

Top Keywords

elekta 75/25
12
linear accelerator
12
influence brass
8
brass compensator
8
compensator thickness
8
thickness field
8
field size
8
size neutron
8
neutron contamination
8
contamination spectrum
8

Similar Publications

Background: Determination of the intradural unruptured paraclinoid aneurysm localization is difficult, but critical for selection of the optimal treatment strategy.

Objective: To assess efficacy of the three-dimensional (3D) evaluation of unruptured paraclinoid aneurysms within Leksell GammaPlan® (LGP; Elekta AB; Stockholm, Sweden) for determination of their intradural localization.

Methods: Overall, 125 incidentally diagnosed unruptured paraclinoid aneurysms in 118 patients (mean age, 55 years) underwent 3D evaluation of their localization within LGP using post-contrast thin-slice constructive interference in steady state (CISS) images, which in 41 cases were additionally co-registered and fused with the axial computed tomography angiography (CTA) source images.

View Article and Find Full Text PDF

Background: One of the most significant Intensity Modulated Radiation Therapy treatment benefits is a high target to normal tissue dose ratio. To improve this advantage, an additional accessory such as a compensator is used to deliver doses. Compensator-based IMRT treatment is usually operated with an energy higher than 10 MV.

View Article and Find Full Text PDF

Radiotherapy represents the most widely spread technique to control and treat cancer. To increase the treatment efficiency, high energy linacs are used. However, applying high energy photon beams leads to a non-negligible dose of neutrons contaminating therapeutic beams.

View Article and Find Full Text PDF

High-energy linear accelerators (linacs) have several advantages, including low skin doses and high dose rates at deep-seated tumours. But, at energies more than 8 MeV, photonuclear reactions produce neutron contamination around the therapeutic beam, which may induce secondary malignancies. In spite of improvements achieved in medical linac designs, many countries still use conventional (non-intensity-modulated radiotherapy) linacs.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!