Article Abstract

Background: Due to the narrow therapeutic range of digoxin, determining serum/plasma digoxin concentrations is critical for assessing patients with congestive heart failure, atrial fibrillation, and certain types of arrhythmias. However, digoxin quantification by competitive immunoassays is susceptible to interferences that may alter the accuracy of its measurement in patient plasma. This study aimed to characterize the extent of bilirubin interference in three commonly used digoxin immunoassays.

Methods: Digoxin concentrations were compared using the Beckman Coulter® Unicel DxI 800, the Vitros® 4600, and the Roche Cobas® 8000 in neat or digoxin-spiked icteric and non-icetric plasma samples. A mixing study was performed to demonstrate how digoxin quantification is affected by bilirubin. An equation was derived that predicts the response of the DxI 800, given known bilirubin and digoxin concentrations.

Results: The DxI reported detectable concentrations of digoxin in high bilirubin samples with no added digoxin, while the Vitros® 4600 and Cobas® 8000 gave virtually undetectable results. Spiking digoxin into samples with elevated bilirubin concentrations resulted in a higher percent recovery for the DxI 800 when compared to the other two platforms. The mixing study also revealed an increase in the percent recovery in the DxI 800, while the Vitros® 4600 and Cobas® 8000 were comparable to the expected concentration of digoxin.

Conclusions: The DxI 800 is most prone to interference by bilirubin, while the Vitros® 4600 and Cobas® 8000 are relatively unaffected. Icteric samples should be interpreted with caution if digoxin quantification is needed, especially on the DxI 800 assay.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2018.10.006DOI Listing

Publication Analysis

Top Keywords

dxi 800
24
vitros® 4600
16
cobas® 8000
16
digoxin
12
digoxin quantification
12
4600 cobas®
12
bilirubin interference
8
interference three
8
three commonly
8
commonly digoxin
8