Introduction: During clinical use, gastrointestinal endoscopes are grossly contaminated with patient's native flora. These endoscopes undergo reprocessing to prevent infectious transmission upon future use. Endoscopy-associated infections and outbreaks have been reported, with a recent focus on the transmission of multi-drug resistant organisms. This review aims to provide an update on endoscopy-associated infections, and the factors contributing to their occurrence.

Methods: PubMed, ScienceDirect, and CINAHL were searched for articles describing gastrointestinal endoscopy-associated infections and outbreaks published from 2008 to 2018. Factors contributing to their occurrence, and the outcomes of each outbreak were also examined.

Results: This review found 18 articles, 16 of which described duodenoscope-associated infections, and the remaining two described colonoscope- and gastroscope-associated infection respectively. Outbreaks were reported from the United States, France, China, Germany, the Netherlands and the United Kingdom. The causative organisms reported were Klebsiella pneumoniae, Pseudomonas aeruginosa, Escherichia coli and Salmonella enteritidis.

Conclusions: A number of factors, including lapses in reprocessing, biofilm formation, endoscope design issues and endoscope damage, contribute to gastrointestinal endoscopy associated infection. Methods of improving endoscope reprocessing, screening for contamination and evaluating endoscope damage may be vital to preventing future infections and outbreaks.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6182826PMC
http://dx.doi.org/10.1186/s12941-018-0289-2DOI Listing

Publication Analysis

Top Keywords

endoscopy-associated infections
16
infections outbreaks
12
gastrointestinal endoscopy-associated
8
outbreaks reported
8
factors contributing
8
endoscope damage
8
infections
6
update gastrointestinal
4
endoscopy-associated
4
infections contributing
4

Similar Publications

The use of reusable flexible endoscopes has increased dramatically over the past decade, however despite improvements in endoscope reprocessing, the continued emergence of endoscopy-associated outbreaks as a result of multi-drug resistant bacteria has highlighted the need for a new approach to disinfection. Here, the use of plasma activated liquids (PALs) for the elimination of mixed species biofilm contamination within the working channels of endoscopes was evaluated. Cold atmospheric pressure plasma was used to chemically activate water and a commercially available pH buffered peracetic acid to create PALs.

View Article and Find Full Text PDF

Comparison of two endoscope channel cleaning approaches to remove cyclic build-up biofilm.

J Hosp Infect

August 2024

Medical Affairs, Nanosonics Ltd, Sydney, Australia. Electronic address:

Introduction: Biofilm contributes significantly to bacterial persistence in endoscope channels. Enhanced cleaning methods capable of removing biofilm from all endoscope channels are required to decrease infection risk to patients. This head-to-head study compared cyclic build-up biofilm removal of an automated endoscope channel cleaner (AECC) with standard manual cleaning according to instructions for use (IFU) in polytetrafluorethylene channels.

View Article and Find Full Text PDF

Background: Current duodenoscope reprocessing protocols are insufficient to prevent contamination and require adaptations to prevent endoscopy-associated infections (EAIs). This study aimed to investigate the effect of a new endoscope cleaning brush on the contamination rate of ready-to-use duodenoscopes.

Methods: This retrospective before-and-after intervention study collected duodenoscope surveillance culture results from March 2018 to June 2022.

View Article and Find Full Text PDF

Background: Qualified reprocessing, of which meticulous channel brushing is the most crucial step, is essential for prevention and control of endoscopy-associated infections. However, channel brushing is often omitted in practice. This study aimed to evaluate the effect of an automated flexible endoscope channel brushing system (AECBS) on improving the quality of endoscope reprocessing.

View Article and Find Full Text PDF
Article Synopsis
  • * Out of 133 sampled endoscopes, 47% and 32% were found contaminated in two phases, with common bacteria including Pseudomonas aeruginosa and Escherichia coli, many of which were multidrug-resistant.
  • * Molecular testing revealed that some endoscopes were contaminated with viruses like hepatitis B, C, and HIV, indicating a significant risk for endoscopy-associated infections and related health complications in the local population.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!