The aim of this study was to evaluate the new commercial Unyvero i60 ITI multiplex PCR system (Curetis, Holzgerlingen, Germany) on native cardiac valves in comparison with made in-house 16S rRNA PCR amplification (91E/13BS primers) and conventional microbiological techniques. Forty-four patients (30 men, 14 women) with suspected infective endocarditis (IE) were included in this evaluation corresponding to 30 aortic valves and 14 mitral valves. IE was definite for 40 patients using the modified Duke criteria. 16S rRNA PCR amplification was successful in 22 patients (55%). The Unyvero i60 ITI cartridge yielded a positive result in 16 patients (40%). Among the 40 cases, the etiological agent was not included in the panel of Unyvero i60 ITI cartridge for 14 cases. Moreover, for S. aureus, the Unyvero i60 ITI cartridge quickly yielded the susceptibility to meticillin. The result of the experiment was available after 5 hours whereas 16S rRNA PCR amplification-sequencing needs 14 hours of manipulation. If the manufacturer incorporates new targets able to detect more endocarditis agents such as viridans streptococci, the Unyvero i60 ITI cartridge may be a promising and easy-to-use test.