Assessment of subjective refraction with a clinical adaptive optics visual simulator.

J Cataract Refract Surg

Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain.

Published: January 2019

Purpose: To clinically validate an adaptive optics visual simulator (VAO) that measures subjective refraction and visual acuity.

Setting: Optics Laboratory, University of Murcia, Murcia, Spain.

Design: Prospective case series.

Methods: Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters: limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC).

Results: Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity).

Conclusion: Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6320260PMC
http://dx.doi.org/10.1016/j.jcrs.2018.08.022DOI Listing

Publication Analysis

Top Keywords

subjective refraction
28
visual simulator
20
adaptive optics
16
optics visual
16
gold standard
16
refraction visual
12
visual
10
visual acuity
8
eyes volunteers
8
interexaminer reproducibility
8

Similar Publications

Purpose: To examine the prevalence and associations of anisometropia with spherical ametropia, cylindrical power, age, and sex.

Methods: Anisometropia was analyzed for subjective refraction. In total, 134,603 refractive surgery candidates were included in the period from 2010 to 2020 at the CARE Vision Refractive Centers in Germany.

View Article and Find Full Text PDF

Use of refractive aids among adults in a general population.

Sci Rep

January 2025

Department of Clinical Medicine, Aarhus University, Aarhus N, DK-8200, Denmark.

Spectacles and contact lenses are important for conducting a normal life in a large part of the general population. The purpose of this study was to estimate the use of these refractive aids in a normal adult population, and to identify characteristics of persons who should be targeted in order to improve uncorrected refraction. In the FORSYN study, 10,350 citizens representative for the adult Danish population were invited for a non-cycloplegic eye examination.

View Article and Find Full Text PDF

Objectives: To compare the efficacy of two different silicone hydrogel bandage contact lenses (BCLs) in terms of visual rehabilitation and ocular discomfort following photorefractive keratectomy (PRK).

Materials And Methods: This prospective study included 60 eyes of 30 patients who underwent bilateral PRK surgery to correct myopia and/or astigmatism refractive errors. Following surgery, lotrafilcon A BCLs were applied to the right eye and senofilcon A BCLs were applied to the left eye.

View Article and Find Full Text PDF

Individual Variations in Vergence and Accommodation Responses Following Virtual Reality Gameplay.

Vision (Basel)

December 2024

Department of Optometry and Vision Science, Faculty of Science and Technology, University of Latvia, LV-1004 Riga, Latvia.

Virtual reality (VR) can challenge the visual system, leading to temporary oculomotor changes, though the degree of change varies among individuals. While the vergence and accommodation system plays a crucial role in VR perception, it remains unclear whether individuals whose visual functions fall outside clinical norms experience larger changes. Thus, our study aimed to investigate whether changes in vergence and accommodation responses following VR gameplay differ between individuals with and without non-strabismic binocular and accommodative disorders.

View Article and Find Full Text PDF

Purpose: To compare three biometers equipped with swept-source optical coherence tomography (OCT), including ARGOS (OCTB1), IOLMaster 700 (OCTB2), and Anterion (AS-OCTB). The primary aim was to assess the axial length (AL) acquisition success rates, and secondary aims included comparing parameters obtained from the three devices and evaluating postoperative refractive prediction errors.

Setting: Chukyo Eye Clinic, Nagoya, Japan.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!