Purpose: In revision hip arthroplasty, custom-made implants are one option in patients with acetabular Paprosky III defects.
Methods: In a retrospective analysis, we identified 11 patients undergoing cup revision using a custom-made implant. The accuracy of the intended position of the implant was assessed on post-operative 3D CT and compared to the pre-operative 3D planning in terms of inclination, anteversion, and centre of rotation. In addition, the accuracy of post-operative plain radiographs for measuring implant position was evaluated in relation to the 3D CT standard.
Results: We found a mean deviation between the planned and the final position of the custom-made acetabular implant on 3D CT of 3.6° ± 2.8° for inclination and of - 1.2° ± 7.0° for anteversion, respectively. Restoration of center of rotation succeeded with an accuracy of 0.3 mm ± 3.9 mm in the mediolateral (x) direction, - 1.1 mm ± 3.8 mm in the anteroposterior (y) direction, and 0.4 mm ± 3.2 mm in the craniocaudal (z) direction. The accuracy of the post-operative plain radiographs in measuring the position of the custom-made implant in relation to 3D CT was 1.1° ± 1.7° for implant inclination, - 2.6° ± 1.3° for anteversion and 1.3 mm ± 3.5 mm in the x-direction, and - 0.9 mm ± 3.8 mm in the z-direction for centre of rotation.
Conclusion: Custom-made acetabular implants can be positioned with good accuracy in Paprosky III defects according to the pre-operative planning. Plain radiographs are adequate for assessing implant position in routine follow-up.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00264-018-4193-3 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!