Background: Interleukin-2 (IL-2) and soluble CD25 (sCD25) are among the most important cytokines and diagnostic biomarkers in hemophagocytic lymphohistiocytosis (HLH). Detecting serum level of IL-2 and sCD25 is valuable for making clinical diagnosis and treatment decision in HLH.

Methods: Since tests showing serum IgG antibody against IL-2 or sCD25 have never been carried out, a new protein biochip, which was modified with cysteine and activated sophorolipid (Cys-SL), was developed.

Results: Limits of detection on the biochip were 78 pg/ml for IL-2 and 39 pg/ml for sCD25, respectively. The data showed that on-chip seroimmunological responses to IL-2 and sCD25 proteins were 20.8% and 83.1% and the seroprevalence of IL-2 and sCD25 IgG antibodies were 45.5% and 57.2%, respectively. Data collection for the seroprevalence of serum antigen-antibody complex of sCD25 was 68.8%. The new biochip model shared similar sensitivity and specificity to chemiluminescent immunoassay (CLIA) in its measuring capacity of serum sCD25.

Conclusions: We addressed and confirmed the involvement of serum IgG antibodies against IL-2 and sCD25 as well as Ag-Ab complex of sCD25 in HLH patients. Therefore, this biochip platform would offer a new technological substitution for clinical serological diagnosis of HLH.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2018.10.008DOI Listing

Publication Analysis

Top Keywords

il-2 scd25
20
serum igg
12
igg antibodies
12
scd25
9
protein biochip
8
il-2
8
antibodies il-2
8
il-2 soluble
8
soluble cd25
8
hemophagocytic lymphohistiocytosis
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!