PATIENT DOSES IN COMMON DIAGNOSTIC X-RAY EXAMINATIONS.

Radiat Prot Dosimetry

Department of Medical Physics, School of Medicine, University of Patras, Patras, Greece.

Published: July 2019

A local survey was conducted, to evaluate the radiation dose to adult patients who underwent diagnostic X-ray examinations. Patient-related and technical data were recorded, in 1504 patients, for each of the 11 individual projections, of the 7 most common examinations performed in an X-ray room, with 1 digital radiography system. The patient entrance surface air kerma (ESAK) and the effective dose (ED) were calculated based on the X-ray tube output and the exposure parameters, as well as utilisation of suitable conversion coefficients, respectively. The 75th percentiles of the distribution of the ESAK and kerma area product (KAP) values were also established. The mean, median and 75th percentiles were compared with the national reference levels and the most common values reported at the European level through the DOSE DATAMED II project. The corresponding ED values were also compared with the average values reported for all European countries. The mean ESAK, KAP and ED values along with the uncertainty U values for chest PA, chest LAT, cranium AP, cranium LAT, cervical spine AP, cervical spine LAT, lumbar spine AP, lumbar spine LAT, pelvis AP, abdomen AP, kidneys and urinary bladder (KUB) AP were 0.12 (0.001) mGy, 0.66 (0.023) mGy, 1.01 (0.034) mGy, 0.69 (0.098) mGy, 0.72 (0.014) mGy, 0.63 (0.011) mGy, 4.12 (0.050) mGy, 5.74 (0.082) mGy, 2.57 (0.024) mGy, 1.94 (0.017) mGy, 2.47 (0.073) mGy, and 0.09 (0.001) Gy cm2, 0.38 (0.012) Gy cm2, 0.32 (0.009) Gy cm2, 0.27 (0.052) Gy cm2, 0.17 (0.004) Gy cm2, 0.21 (0.006) Gy cm2, 1.18 (0.018) Gy cm2, 1.86 (0.023) Gy cm2, 1.41 (0.012) Gy cm2, 1.27 (0.010) Gy cm2, 1.28 (0.038) Gy cm2, as well as 0.01 (0.0001) mSv, 0.05 (0.0016) mSv, 0.02 (0.0006) mSv, 0.01 (0.0012) mSv, 0.03 (0.0008) mSv, 0.03 (0.0006) mSv, 0.26 (0.0038) mSv, 0.17 (0.0022) mSv, 0.20 (0.0016) mSv, 0.23 (0.0018) mSv, 0.23 (0.0068) mSv, respectively. The 75th percentiles along with the uncertainty U values for chest PA, chest LAT, cranium AP, cranium LAT, cervical spine AP, cervical spine LAT, lumbar spine AP, lumbar spine LAT, pelvis AP, abdomen AP, kidneys and urinary bladder (KUB) AP were 0.14 (0.006) mGy, 0.88 (0.031) mGy, 1.22 (0.049) mGy, 0.94 (0.098) mGy, 0.93 (0.027) mGy, 0.78 (0.013) mGy, 5.16 (0.073) mGy, 7.24 (0.134) mGy, 2.96 (0.047) mGy, 2.59 (0.036) mGy, 3.07 (0.116) mGy, as well as 0.10 (0.0006) Gy cm2, 0.51 (0.017) Gy cm2, 0.37 (0.020) Gy cm2, 0.33 (0.040) Gy cm2, 0.23 (0.007) Gy cm2, 0.26 (0.011) Gy cm2, 1.50 (0.036) Gy cm2, 2.26 (0.035) Gy cm2, 1.61 (0.023) Gy cm2, 1.67 (0.017) Gy cm2, 1.56 (0.069) Gy cm2, in terms of ESAK and KAP values, respectively. The results were significantly lower compared with the national reference levels, the most common DRL values reported at the European level and other previously reported dose values. Patient dose surveys could contribute towards optimising radiation protection for patients, therefore, highlighting the necessity to increase the awareness and knowledge of the radiation dose in conjunction with the required image quality.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncy169DOI Listing

Publication Analysis

Top Keywords

mgy
22
cm2
22
cervical spine
16
spine lat
16
lumbar spine
16
75th percentiles
12
kap values
12
values reported
12
reported european
12
msv
11

Similar Publications

Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).

Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.

View Article and Find Full Text PDF

Purpose: To evaluate and compare lung nodules' image quality and radiation dose exposure using photon-counting computed tomography (PC-CT) and conventional energy-integrating detector computed tomography (EID-CT).

Methods: Protocol pre-registration was performed a priori at (https://osf.io/krj5y/).

View Article and Find Full Text PDF

Omnipolar mapping versus point-by-point mapping approach for catheter ablation of atrioventricular accessory pathway.

J Interv Card Electrophysiol

January 2025

Department of Cardiovascular Medicine, Dokkyo Medical University, 880 Kitakobayashi, Mibu, Shimotsugagun, Tochigi, 321-0293, Japan.

Background: The conventional mapping approach for the atrioventricular accessory pathway (AP) involves point-by-point mapping to identify the connection sites of the AP to the atria or ventricle and accurate interpretation of local electrograms. Omnipolar mapping technology (OMT) explains how vector and wave speed are produced by using both unipolar and bipolar signals to obtain omnipolar signals, directions, and conduction velocity. The aim of this study is to verify the effectiveness of OMT for catheter ablation of AP.

View Article and Find Full Text PDF

Background: Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.

Purpose: To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.

View Article and Find Full Text PDF

Introduction: Radiation exposure during percutaneous nephrolithotomy (PCNL) can vary depending on the method used for renal access. This study aimed to compare fluoroscopy time and dose during PCNL when renal access is achieved via interventional radiology (IR) versus urology.

Methods: A retrospective review of patients who underwent unilateral PCNL between January 2020 and February 2023 was conducted.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!