Introduction: Reduction of radiation dosage in the pediatric cardiac catheterization laboratory (PCL) is important to reduce the risk of its stochastic effect in children with congenital heart disease. Lowering the frame rate would reduce radiation dosage possibly at the expense of image quality, potentially resulting in higher fluoroscopic time and procedural complication rate.

Methods: The data were retrospectively analyzed in three eras: era 1 (n = 234), cineangiography 30 frames/sec (f/s) and fluoroscopy 15 pulse/sec (p/s); era 2 (n = 381), cineangiography 30 f/s and fluoroscopy 6 p/s; and era 3 (n = 328), cineangiography 15 f/s and fluoroscopy 6 p/s. Also, three operators blinded to the frame rate setting evaluated cineangiography image quality. In this study, the impact of lowering the default frame rates on radiation dosage, fluoroscopic time, contrast volume, diagnostic image quality, and complication rates in the PCL was assessed.

Results: Overall radiation dosage progressively declined during these eras (70.0 vs 64.1 vs 36.6 µGym /kg, P < .001) without a difference in significant adverse event rates. There was no significant increase in either fluoroscopy time or contrast volume. There was no difference in the diagnostic image quality between cineangiography 30 and 15 f/s. Lowering the default frame/pulse rates of both fluoroscopy and cineangiography significantly decreased the overall radiation dosage in the PCL. Importantly, fluoroscopy time, contrast volume, and complication rates did not increase while maintaining diagnostic image quality.

Conclusion: This quality improvement project proved successful in lowering radiation dosage without compromising the efficacy and safety of catheterizations.

Download full-text PDF

Source
http://dx.doi.org/10.1111/chd.12677DOI Listing

Publication Analysis

Top Keywords

radiation dosage
16
frame rate
12
image quality
12
lowering frame
8
cardiac catheterization
8
fluoroscopic time
8
fluoroscopy 6 p/s
8
reducing radiation
4
radiation exposure
4
exposure lowering
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!