Electron cyclotron emission imaging (ECEI) and microwave imaging reflectometry diagnostics have been employed on a number of magnetic fusion plasma confinement devices. The common approach is based on a Gaussian beam assumption, which generates good spatial resolution (centimeter level). However, the radial focal depth is limited by the poloidal resolution, which is comparable with the Rayleigh length (∼150 mm). By contrast, a new Bessel beam approach has been developed and demonstrated to generate much longer focal depth with the property of propagation stability. To test the new approach, the DIII-D tokamak LCP ECEI optics have been re-designed to support a Bessel beam approach based on an axicon lens. The achievable radial coverage can exceed that of the current Gaussian approach by 3×. The imaging result is discussed in this paper based on the simulation analysis and laboratory testing result.