Purpose: The aim of this study was to compare the usability and patient-rated experiences of an autoinjector with a prefilled syringe in patients with migraine, who self-administered galcanezumab, and to compare pharmacokinetic parameters between these devices.
Materials And Methods: Patient-rated experiences with an investigational autoinjector and a prefilled syringe were compared in an open-label, 12-month study of once-monthly injections of galcanezumab 120 or 240 mg (NCT02614287). Patient-rated ease of usability was assessed with the Subcutaneous Administration Assessment Questionnaire (SQAAQ) and compared between devices. Positive responses on the SQAAQ were rated as "agree or strongly agree" to 12 statements. Tolerability was assessed by the frequency of injection-site-related adverse events (AEs) by device and injection location. In a separate study, galcanezumab pharmacokinetics in healthy subjects was compared between the devices (NCT02836613).
Results: In the open-label clinical trial, 179 patients used both the prefilled syringe and autoinjector at least once. The majority of patients (91%-97%) had positive responses on the SQAAQ to the use of autoinjector across the items assessed. There were 23 injection-site-related AEs with the first self-administered injection with the prefilled syringe (N=7) or autoinjector (N=16; =0.061), with the most common AE for either device being injection-site pain. There were no significant between-device differences in injection-site-related AEs. For pharmacokinetics, the 90% CI for the ratio (autoinjector/prefilled syringe) of geometric least-square means for the galcanezumab area under the curve (AUC) concentration and maximum concentration () was between 0.8 and 1.25, indicating no statistically significant difference in the galcanezumab concentrations regardless of the device used.
Conclusion: The ease of usability with either device was comparable, and there were no significant differences in tolerability between the prefilled syringe and autoinjector with the first self-administration; however, the analysis was not powered to detect a clinically significant difference. Galcanezumab pharmacokinetics were comparable between devices.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6147689 | PMC |
http://dx.doi.org/10.2147/PPA.S170636 | DOI Listing |
Expert Rev Pharmacoecon Outcomes Res
January 2025
Merck & Co. Inc, Rahway, NJ, USA.
Background: We evaluated UK nurses' preferences for pediatric hexavalent vaccine attributes.
Research Design And Methods: In a discrete-choice experiment study, 150 nurses chose between 2 hypothetical pediatric hexavalent vaccines with varying attribute levels (device type, plastic in packaging, time on the market, and time the vaccine can stay safely at room temperature) in a series of choice questions. Using random-parameters logit-model estimates, conditional relative attribute importance (CRAI) and odds ratios (ORs) were calculated.
Patient
January 2025
Division of Rheumatology, Allergy and Immunology, Department of Medicine, University of Cincinnati College of Medicine, Cincinnati, OH, USA.
Background: In the context of injectable biologic products approved or in development for chronic spontaneous urticaria (CSU), it is important to capture which treatment attributes matter most to patient and what trade-offs patients are willing to make.
Objectives: The CHOICE-CSU study aimed to quantify patient preferences toward injectable treatment attributes among patients with CSU, inadequately controlled by H1-antihistamines.
Methods: This was a two-phase cross-sectional patient preference study in adult patients with a diagnosis of CSU, inadequately controlled by H1-antihistamines.
Eur J Hosp Pharm
January 2025
Pharmacy Department, Lausanne University Hospital, Lausanne, Switzerland
Background: Intravitreal and intracameral administration of melphalan and topotecan (TPT) has shown its efficacy in the treatment of retinoblastoma over the last few years. Due to rapid hydrolysis, melphalan must be administered within the hour following reconstitution. With improved stability at room temperature and reduced ocular toxicity, TPT seems to be a promising candidate for production of prefilled syringes in terms of safety and efficiency of preparation and treatment administration.
View Article and Find Full Text PDFOphthalmologie
January 2025
Augenklinik Petrisberg, Max-Planck-Str. 14-16, 54296, Trier, Deutschland.
Expert Rev Clin Immunol
January 2025
Division of Rheumatology, Department of Medicine III, Medical University of Vienna, Vienna, Austria.
Background: CT-P47 is a candidate tocilizumab biosimilar that is currently in clinical development. We assessed the usability of CT-P47 self-administration via auto-injector (AI) in patients with rheumatoid arthritis (RA).
Research Design And Methods: This was a 12-week, single-arm, open-label, multiple-dose, Phase 3 study.