Purpose: Single radius knee implants were introduced to reduce the level of paradoxical anterior femoral translation (AFT) during mid-flexion after total knee arthroplasty. Findings from clinical and experiment studies are inconsistent, which may be due to the different loading conditions and articular conformities of the knee implants studied. The aim of this study is to analyze how variations in these two factors affect the mid-flexion stability of a single radius knee prosthesis.
Methods: Six daily activities (walking, stair ascent, stair descent, sit-to-stand, pivot turn and crossover turn), and three articular conformity ratios (low, moderate and high) were considered. The resulting AFTs from the 18 finite element models were analyzed.
Results: For low conformity knees, the worst case activity (the greatest AFT) was sit-to-stand with an AFT of 6.2 mm, while for the moderate conformity the worst case was crossover turn and pivot turn. For high conformity, all activities produced a relatively small AFT ranging from 0 mm to 1.8 mm, which more closely resembles natural knee motion. In addition, no AFT was recorded during stair ascent for all three conformities (low, moderate, high).
Conclusions: This study demonstrated that the amount of AFT is highly dependent on the activity being undertaken and the articular conformity of the knee prosthesis, and the worst case activity depends on the knee conformity. The clinical relevance of this study is that it offers valuable information towards the design of improved knee prostheses and selection of knee implants for clinical use.
Level Of Evidence: II.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00167-018-5181-2 | DOI Listing |
Biomed Mater
January 2025
School of Advanced Manufacturing, Nanchang University - Qianhu Campus, Nanchang, Jiangxi, China, Nanchang, --- Select One ---, 330031, CHINA.
The articular cartilage is characterized by its gradient hierarchical structure, which exhibits excellent lubrication and robust load-bearing properties. However, its inherent difficulty in self-repair after damage presents numerous formidable challenges for cartilage repair. Inspired by the unique structure of articular cartilage, a biomimetic bilayer hydrogel composed of PAM (polyacrylamide) and PAM/SA (sodium alginate) is prepared using a two-step in-situ swelling method.
View Article and Find Full Text PDFBMC Musculoskelet Disord
December 2024
The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei, 050051, P.R. China.
Background: Limitations in forearm rotation resulting from distal radius fracture are often neglected in clinical practice. We aimed to explore possible influencing factors of forearm rotation limitation following conservative treatment of these fractures.
Methods: A series of patients with distal radius fractures who underwent conservative treatment in the Third Hospital of Hebei Medical University were retrospectively enrolled.
Acta Orthop
December 2024
Department of Biomedical Engineering, University of California, Davis, CA; Department of Mechanical Engineering, University of California, Davis, CA; Department of Orthopaedic Surgery, University of California Davis Medical Center, Sacramento, CA, USA.
Background And Purpose: In total knee arthroplasty (TKA), an insert with ball-in-socket (BS) medial conformity (MC) and posterior cruciate ligament (PCL) retention restores kinematics closer to native than an insert with intermediate (I) MC. However, high medial conformity might compromise baseplate stability as indicated by maximum total point motion (MTPM). Using the BS MC insert with PCL retention, we aimed to determine whether (i) the baseplate is stable as indicated by mean MTPM < 0.
View Article and Find Full Text PDFBMC Musculoskelet Disord
December 2024
Department of Trauma Orthopaedics and Hand Surgery, The Fifth Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, China.
Background: Sternoclavicular joint dislocation (SJD) is rare, and joint instability after dislocation easily leads to deformity, pain, and limitations in performing activities, often requiring surgical treatment. Currently, there is no ideal internal fixation method for SJD.
Case Presentation: We report the case of a 38-year-old female patient with anterior dislocation of the right sternoclavicular joint (SJ) caused by a car accident who underwent open reduction and internal fixation using the double plate technique combined with the cable technique.
Hepatol Int
December 2024
Division of Hepatology, Department of Gastroenterology and Hepatology, Chang Gung Memorial Hospital at Linkou, No 5, Fu Hsing Street, Kuei Shan, Taoyuan, Taiwan.
Background: The phenotype of cryoglobulinemia in hepatitis B virus (HBV) carriers remains elusive.
Methods: A 7-year prospective cohort of 648 hepatitis B e antigen (HBeAg)-negative Taiwanese HBV carriers [males: 344 (53%)] was conducted.
Results: Among 648, 189 (29.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!