Optimization of Electrospray Ionization Source Parameters for Lipidomics To Reduce Misannotation of In-Source Fragments as Precursor Ions.

Anal Chem

Department of Medicine , Division of Sleep and Circadian Disorders, Brigham and Women's Hospital, and Department of Medicine, Division of Sleep Medicine, Harvard Medical School, Boston , Massachusetts 02115 , United States.

Published: November 2018

AI Article Synopsis

Article Abstract

Lipidomics requires the accurate annotation of lipids in complex samples to enable determination of their biological relevance. We demonstrate that unintentional in-source fragmentation (ISF, common in lipidomics) generates ions that have identical masses to other lipids. Lysophosphatidylcholines (LPC), for example, generate in-source fragments with the same mass as free fatty acids and lysophosphatidylethanolamines (LPE). The misannotation of in-source fragments as true lipids is particularly insidious in complex matrixes since most masses are initially unannotated and comprehensive lipid standards are unavailable. Indeed, we show such LPE/LPC misannotations are incorporated in the data submitted to the National Institute of Standards and Technology (NIST) interlaboratory comparison exercise. Computer simulations exhaustively identified potential misannotations. The selection of in-source fragments of highly abundant lipids as features, instead of the correct recognition of trace lipids, can potentially lead to (i) missing the biologically relevant lipids (i.e., a false negative) and/or (ii) incorrect assignation of a phenotype to an incorrect lipid (i.e., false positive). When ISF is not eliminated in the negative ion mode, ∼40% of the 100 most abundant masses corresponding to unique phospholipids measured in plasma were artifacts from ISF. We show that chromatographic separation and ion intensity considerations assist in distinguishing precursor ions from in-source fragments, suggesting ISF may be especially problematic when complex samples are analyzed via shotgun lipidomics. We also conduct a systematic evaluation of electrospray ionization (ESI) source parameters on an Exactive equipped with a heated electrospray ionization (HESI-II) source with the objective of obtaining uniformly appropriate source conditions for a wide range of lipids, while, at the same time, reducing in-source fragmentation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6297073PMC
http://dx.doi.org/10.1021/acs.analchem.8b03436DOI Listing

Publication Analysis

Top Keywords

in-source fragments
20
electrospray ionization
12
source parameters
8
misannotation in-source
8
precursor ions
8
complex samples
8
in-source fragmentation
8
in-source
7
lipids
7
fragments
5

Similar Publications

Peptide mapping analysis of synthetic semaglutide and liraglutide for generic development of drugs originating from recombinant DNA technology.

J Pharm Biomed Anal

January 2025

College of Pharmacy, Chung-Ang University, Seoul 06974, Republic of Korea; Department of Global Innovative Drugs, The Graduate School of Chung-Ang University, Seoul 06974, Republic of Korea. Electronic address:

Semaglutide and liraglutide are long-acting glucagon-like peptide-1 receptor agonists used to treat type-2 diabetes and obesity. Recent advances in peptide synthesis and analytical technologies have enabled the development of synthetic generic peptide for reference listed drugs (RLD) originating from recombinant DNA (rDNA) technology. Since the original semaglutide and liraglutide were produced through rDNA technology, there has been great interest in developing their synthetic peptides as generic versions of the original drugs.

View Article and Find Full Text PDF

RIBULOSE-1,5-BISPHOSPHATE CARBOXYLASE/OXYGENASE (RUBISCO) is the most abundant enzyme and CO2 bio-sequestration system on Earth. Its in vivo activity is usually determined by 14CO2 incorporation into 3-phosphoglycerate (3PGA). However, the radiometric analysis of 3PGA does not distinguish carbon positions.

View Article and Find Full Text PDF

mpactR automates pre-processing of liquid chromatography-tandem mass spectrometry (LC-MS/MS) data from microbiological samples to correct mispicked peaks, resolve inter-sample variation in abundance across technical replicates, account for in-source ion fragmentation, and remove background noise to yield high-quality mass spectrometry features. The package is available through CRAN and GitHub.

View Article and Find Full Text PDF

Improving the reliability of phthalate esters analysis in water samples by gas chromatography-tube plasma ionization-high-resolution mass spectrometry (GC-TPI-HRMS).

Talanta

April 2025

Applied Analytical Chemistry, University of Duisburg-Essen, Universitatsstr. 5, 45141, Essen, Germany; Teaching and Research Center for Separation, University of Duisburg-Essen, Universitatsstr. 5, 45141, Essen, Germany. Electronic address:

The monitoring of phthalate esters (PAEs) is challenging due to background contamination as well as the low selectivity observed when analyzing them by gas chromatography coupled to mass spectrometry (GC-MS) using electron ionization (EI). In this sense, alternative and soft ionization techniques could help to enhance the performance of the analytical determinations of PAEs in food samples. In this work, the use of a novel and soft ionization technique tube plasma ionization (TPI) has been explored to enhance the selectivity and sensitivity in the determination of PAEs in drinking water samples with GC-MS.

View Article and Find Full Text PDF

Introduction: Liquid chromatography-mass spectrometry (LC-MS) has enhanced the rapid, accurate analysis of complex plant extracts, eliminating the need for extensive isolation. Tandem mass spectrometry (MS/MS) further enhances this process by providing detailed structural information. However, differentiating structural isomers remains a challenge due to their minor spectral and structural differences.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!