Background: Arrhythmias in children are often paroxysmal, complicating the ability to capture the abnormal rhythm on routine ECG during an outpatient visit. The Alivecor Kardia Mobile (KM) device is a wireless mobile health (mHealth) device that generates a single lead ECG tracing with a FDA-approved algorithm for detection of atrial fibrillation in adults.

Objective: The goal of this study is to assess the accuracy of interval measurements on KM tracings by directly comparing to standard 12-lead ECGs in pediatric patients.

Methods: This single center, prospective study enrolled pediatric outpatients, age <18 years presenting for cardiology clinic visits, into 3 groups based on age: 0-5 years, 6-10 years, and 11-18 years. Patients were excluded if 12-lead ECG was not ordered during the visit. Each enrolled subject underwent standard 12-lead ECG followed by 30-second KM tracing. ECG parameters were batch read by 2 blinded pediatric electrophysiologists.

Results: Thirty patients were recruited with 10 patients/group. Structural heart disease and/or conduction abnormality was present in 20 patients (67%). Majority of tracings (27/30, 90%) were of diagnostic quality on first attempt. Overall, the ΔPR was 15.2±10.8ms (r = 0.86), ΔQRS was 9.6±8ms (r = 0.86), and ΔQTc was 15.6±12.7ms (r = 0.83). There were 9 patients with ΔQTc measurements >20ms with 4/9 (44%) having a conduction disorder and 2/9 (22%) having marked sinus arrhythmia. Bland-Altman method of agreement demonstrated strong agreement for QRSd and QTc. The AF algorithm reported 4/30 (13%) false positive "possible AF" diagnoses (rhythm over-read on KM demonstrated n = 3 marked sinus arrhythmia, n = 1 sinus rhythm with aberrated PACs) resulting in a specificity of 87%.

Conclusion: The Alivecor Kardia device produces accurate single lead ECG tracings in both healthy children and children with cardiac disease or rhythm abnormalities across the pediatric spectrum. This mHealth application provides an accurate, non-invasive, real-time approach for ambulatory ECG monitoring in children and adolescents.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6160047PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0204403PLOS

Publication Analysis

Top Keywords

mobile health
8
standard 12-lead
8
alivecor kardia
8
single lead
8
lead ecg
8
marked sinus
8
sinus arrhythmia
8
ecg
6
smartphone wireless
4
wireless ecgs
4

Similar Publications

While telegenetic counseling has increased substantially since the start of the COVID-19 pandemic, previous studies reported concerns around building rapport, nonverbal communication, and the patient-counselor relationship. This qualitative evaluation elicited feedback from genetic counselors, referring clinicians, and patients from a single healthcare organization to understand the user-driven reasons for overall satisfaction and experience. We conducted 22 in-depth, semi-structured interviews with participants from all 3 groups between February 2022 and February 2023.

View Article and Find Full Text PDF

Background: There remains a significant gap in systematic research on healthcare utilization behaviors and the influencing factors for patients with Parkinson's disease (PD), particularly those in late stages.

Methods: PD patients in late stage (Hoehn and Yahr (HY) stages 4 and 5) and their caregivers from Seoul National University Hospital Movement Disorders Clinic participated. A total of 103 respondents completed a questionnaire covering medical utilization behaviors, perceptions of face-to-face and telemedicine consultations, and additional medical service needs.

View Article and Find Full Text PDF

Introdution: Although oral and dental diseases may occur in unexpected or even emergency conditions, but some of the diagnosis and treatments can be algorithmically done following some guidelines. The development and implementation of a system that provides users with a record of history and a proposal of required actions can be not only efficiently practical, but also virtually simple.

Materials And Method: A system made up of web and mobile apps is proposed and evaluated for screening and self-care of oral and dental problems and for providing advice on dental emergencies and therapeutic measures.

View Article and Find Full Text PDF

Background: The plethora of troublesome signs and symptoms of multiple sclerosis (MS) reduces patients' quality of life (QOL) and coping skills. Face-to-face (F2F) education is effective and practical as it allows for more engagement and active learning. The use of mobile health technology to enhance health is now an excellent potential to establish a more efficient health system.

View Article and Find Full Text PDF

In this viewpoint, we explore Vietnam's health system vulnerabilities and its national response to the COVID-19 pandemic, as well as critical areas of health system resilience, including health financing, workforce distribution, information systems, and governance. While Vietnam achieved early success through strong governance and mass vaccination campaigns, the pandemic revealed weaknesses in resource procurement, workforce imbalance, and limitations of its health information system. There are challenges in ensuring the rapid disbursement of financial resources and reliance on imported medical supplies, which delayed response times.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!