What We Already Know About This Topic: WHAT THIS ARTICLE TELLS US THAT IS NEW: BACKGROUND:: Intraoperative lung-protective ventilation has been recommended to reduce postoperative pulmonary complications after abdominal surgery. Although the protective role of a more physiologic tidal volume has been established, the added protection afforded by positive end-expiratory pressure (PEEP) remains uncertain. The authors hypothesized that a low fixed PEEP might not fit all patients and that an individually titrated PEEP during anesthesia might improve lung function during and after surgery.
Methods: Forty patients were studied in the operating room (20 laparoscopic and 20 open-abdominal). They underwent elective abdominal surgery and were randomized to institutional PEEP (4 cm H2O) or electrical impedance tomography-guided PEEP (applied after recruitment maneuvers and targeted at minimizing lung collapse and hyperdistension, simultaneously). Patients were extubated without changing selected PEEP or fractional inspired oxygen tension while under anesthesia and submitted to chest computed tomography after extubation. Our primary goal was to individually identify the electrical impedance tomography-guided PEEP value producing the best compromise of lung collapse and hyperdistention.
Results: Electrical impedance tomography-guided PEEP varied markedly across individuals (median, 12 cm H2O; range, 6 to 16 cm H2O; 95% CI, 10-14). Compared with PEEP of 4 cm H2O, patients randomized to the electrical impedance tomography-guided strategy had less postoperative atelectasis (6.2 ± 4.1 vs. 10.8 ± 7.1% of lung tissue mass; P = 0.017) and lower intraoperative driving pressures (mean values during surgery of 8.0 ± 1.7 vs. 11.6 ± 3.8 cm H2O; P < 0.001). The electrical impedance tomography-guided PEEP arm had higher intraoperative oxygenation (435 ± 62 vs. 266 ± 76 mmHg for laparoscopic group; P < 0.001), while presenting equivalent hemodynamics (mean arterial pressure during surgery of 80 ± 14 vs. 78 ± 15 mmHg; P = 0.821).
Conclusions: PEEP requirements vary widely among patients receiving protective tidal volumes during anesthesia for abdominal surgery. Individualized PEEP settings could reduce postoperative atelectasis (measured by computed tomography) while improving intraoperative oxygenation and driving pressures, causing minimum side effects.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/ALN.0000000000002435 | DOI Listing |
Sci Rep
January 2025
IRCCS Fondazione Don Carlo Gnocchi ONLUS, 50143, Florence, Italy.
Bioelectrical Impedance Vector Analysis (BIVA) is a valuable tool for evaluating hydration and body composition, but its application in subacute post-stroke patients remains unexplored. This study aimed to fill this gap by analyzing BIVA in a cohort of 87 subacute post-stroke patients (42 women, mean age 69 ± 12) undergoing rehabilitation. At admission (T0), diagnosis of malnutrition with GLIM criteria and of sarcopenia with EWGSOP2 was done, and patients were analyzed with BIVA.
View Article and Find Full Text PDFCurr Obes Rep
January 2025
Metabolism and Body Composition, Pennington Biomedical Research Center, Baton Rouge, LA, 70808, USA.
Background: Recent technological advances have introduced novel methods for measuring body composition, each with unique benefits and limitations. The choice of method often depends on the trade-offs between accuracy, cost, participant burden, and the ability to measure specific body composition compartments.
Objective: To review the considerations of cost, accuracy, portability, and participant burden in reference and emerging body composition assessment methods, and to evaluate their clinical applicability.
Sensors (Basel)
December 2024
Institute of Robotics, Autonomous System and Sensing, School of Electronic and Electrical Engineering, University of Leeds, Leeds LS2 9JT, UK.
Knee joint disorders pose a significant and growing challenge to global healthcare systems. Recent advancements in robotics, sensing technologies, and artificial intelligence have driven the development of robot-assisted therapies, reducing the physical burden on therapists and improving rehabilitation outcomes. This study presents a novel knee exoskeleton designed for safe and adaptive rehabilitation, specifically targeting bed-bound stroke patients to enable early intervention.
View Article and Find Full Text PDFMolecules
December 2024
Ruđer Bošković Institute, Bijenička Cesta 54, 10000 Zagreb, Croatia.
This study examines the influence of ligand design on the structural, optical, and electrical properties of copper-based coordination complexes. Ligands HL and HL were synthesized via the reaction of 5-nitrosalicylaldehyde with 2-hydroxy- or 4-hydroxybenzhydrazide. HL was obtained from the reaction of carbohydrazide and salicylaldehyde, while HL was prepared by condensing 4-methoxysalicylaldehyde with thiocarbohydrazide.
View Article and Find Full Text PDFCrit Care
January 2025
Departamento de Medicina, Hospital Clínico Universidad de Chile, Unidad de Pacientes Críticos, Dr. Carlos Lorca Tobar 999, Independencia, Santiago, Chile.
Background: Double cycling with breath-stacking (DC/BS) during controlled mechanical ventilation is considered potentially injurious, reflecting a high respiratory drive. During partial ventilatory support, its occurrence might be attributable to physiological variability of breathing patterns, reflecting the response of the mode without carrying specific risks.
Methods: This secondary analysis of a crossover study evaluated DC/BS events in hypoxemic patients resuming spontaneous breathing in cross-over under neurally adjusted ventilatory assist (NAVA), proportional assist ventilation (PAV +), and pressure support ventilation (PSV).