Technical Note: Experimental verification of EGSnrc calculated depth dose within a parallel magnetic field in a lung phantom.

Med Phys

Department of Oncology, Medical Physics Division, University of Alberta, 11560 University Avenue, Edmonton, Alberta, T6G 1Z2, Canada.

Published: December 2018

Purpose: The calculation of depth doses from a 6 MV photon beam in polystyrene using EGSnrc Monte Carlo, within a parallel magnetic field, has been previously verified against measured data. The current work experimentally investigates the accuracy of EGSnrc calculated depth doses in lung within the same parallel magnetic field.

Methods: Two cylindrical bore electromagnets produced a magnetic field parallel to the central axis of a Varian Silhouette beam. A Gammex lung phantom was used, along with a parallel plate ion chamber, for the depth dose measurements. Two experimental setups were investigated: top of phantom coinciding with the top of the magnet's bore, and top of phantom coinciding with the center of the bore. EGSnrc was modified to read the 3D magnetic field distribution and then used to simulate the depth dose in lung.

Results: The parallel magnetic field caused measurable increases in dose at the surface and in the buildup region for both setups. For the setup where the top of the lung phantom coincides with the top of the magnet, the surface dose increased by ~11% compared to the no magnetic field case but the depth of maximum dose remained unchanged. When the phantom's top surface coincided with the center of the magnet, the surface dose increased by 32% and dose maximum occurred at a shallower depth. EGSnrc was able to calculate these dose increases due to the magnetic field accurately for both setups. All the simulated depth dose values were within 2% (with respect to D ) of the measured ones, and most of the investigated points were within 1.5%.

Conclusions: Surface and dose increases due to a parallel magnetic field have been measured in a lung phantom at two separate locations within the magnetic field. EGSnrc has been shown to match these measurements to within 2%.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13215DOI Listing

Publication Analysis

Top Keywords

magnetic field
36
parallel magnetic
20
depth dose
16
lung phantom
16
surface dose
12
dose
11
magnetic
10
field
9
egsnrc calculated
8
depth
8

Similar Publications

Background: In arrhythmogenic cardiomyopathy (ACM), left ventricle-dominant presentation has poorer outcomes than right-dominant presentation, suggesting that interventricular functional disparity might play a role in patients' prognosis. However, the prognostic impact of ventricular functional discordance in ACM patients remains unknown.

Purpose: To assess whether ventricular functional disparity measured as ventricular discordance index, defined as the ratio of right-ventricular ejection fraction (RVEF) to left-ventricular ejection fraction (LVEF), might reveal prognostic disparities between phenotypes and offer added risk stratification value.

View Article and Find Full Text PDF

Magnetic-proximity-induced anomalous hall effect at the EuO/SbTeinterface.

J Phys Condens Matter

December 2024

Department of Physics, Indian Institute of Technology Delhi, DEPRTMENT OF PHYSICS, IIT DELHI, HAUZ KHAS, New Delhi, Delhi, 110016, INDIA.

Time-reversal symmetry breaking of a topological insulator phase generates zero-field edge modes which are the hallmark of the quantum anomalous Hall effect (QAHE) and of possible value for dissipation-free switching or non-reciprocal microwave devices. But present material systems exhibiting the QAHE, such as magnetically doped bismuth telluride and twisted bilayer graphene, are intrinsically unstable, limiting their scalability. A pristine magnetic oxide at the surface of a TI would leave the TI structure intact and stabilize the TI surface, but epitaxy of an oxide on the lower-melting-point chalcogenide presents a particular challenge.

View Article and Find Full Text PDF

Lithium-sulfur batteries (LSBs) face challenges from the shuttle effect of lithium polysulfides (LiPSs) and slow redox kinetics. In this study, a NiCo-Doped 3D Ordered Mesoporous Carbon (NiCo-3DOMC) composite material is synthesized using a gel-crystalline template and sol-gel method to modify polypropylene separators in LSBs. Density Functional Theory calculations and experiment results demonstrate that under a magnetic field, the NiCo-3DOMC enhances adsorption and catalyzes the conversion of LiPSs, effectively mitigating the shuttle effect and boosting redox kinetics.

View Article and Find Full Text PDF

Moiré magnetism and moiré excitons in twisted CrSBr bilayers.

Proc Natl Acad Sci U S A

January 2025

Department of Physics and Astronomy, California State University Northridge, Northridge, CA 91330-8268.

Moiré excitons and moiré magnetism are essential to semiconducting van der Waals magnets. In this work, we perform a comprehensive first-principles study to elucidate the interplay of electronic excitation and magnetism in twisted magnetic CrSBr bilayers. We predict a twist-induced quantum phase transition for interlayer magnetic coupling and estimate the critical twist angle below which moiré magnetism with mixed ferromagnetic and antiferromagnetic domains could emerge.

View Article and Find Full Text PDF

Dendritic cells (DCs) play a crucial role in initiating antitumor immune responses. However, in the tumor environment, dendritic cells often exhibit impaired antigen presentation and adopt an immunosuppressive phenotype, which hinders their function and reduces their ability to efficiently present antigens. Here, a dual catalytic oxide nanosponge (DON) doubling as a remotely boosted catalyst and an inducer of programming DCs to program immune therapy is reported.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!