Purpose: Surgical reconstructions in three dimensions are needed for treatment of foot and ankle deformities. However, surgical results might be influenced by the skill and experience of doctors which complement the limited information for reconstructions in three dimensions. To solve these, studies were carried out to measure plantar pressure distribution during surgery. Though, it was impossible to accurately measure plantar pressure distribution accurately during operation. Therefore, we proposed an intraoperative plantar pressure measurement (IPPM) device that enables proper navigation in the push direction.

Methods: For this purpose, first, we investigated how the physiological load axis passes through the human body to identify the pushing direction of the pressure sensor of the device toward the patient's foot. In particular, we hypothesized that the physiological load axis passes through the femoral head center and we evaluated this in a measurement experiment with nine healthy subjects. Second, based on these results, we developed the IPPM device that has two force sensors to identify the pushing direction toward the femoral head center and a conductive ink sensor to measure plantar pressure distribution. Finally, we conducted the experiments with nine healthy subjects and two users.

Results: From the first experimental results, the physiological load axis was found to pass through the femoral head center in normal standing posture. From the evaluation experiment, there are no significant differences statistically in plantar pressure distributions between the conditions of using IPPM device and without using it for both a medical student and a surgeon. However, in some cases the plantar pressure distribution can be reproduced similarly to that of the standing posture, and also from the evaluation experiment concerning the relation between CoP position and NCC, the NCC tends to increase when the position of the CoP is closer to that at the standing posture.

Conclusion: The IPPM device has possibility to reproduce the plantar pressure distribution during surgery and prevent the recurrence of surgical complications.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11548-018-1862-zDOI Listing

Publication Analysis

Top Keywords

plantar pressure
32
pressure distribution
20
ippm device
16
measure plantar
12
physiological load
12
load axis
12
femoral head
12
head center
12
pressure
9
plantar
8

Similar Publications

This contribution details a new high-fidelity finite element analysis (FEA) methodology for the investigation of the effect of the graft size on the pressure distribution developing at the calcaneocuboid joint after the Evans osteotomy procedure. The FEA model includes all 28 bones of the foot up to the distal end of fibula and tibia as well as soft tissues, tendons, and muscles. The developed FEA model was validated by comparing the in-vivo pressure distribution on the foot plantar with the in-silico results, resulting in a low deviation equal to 7.

View Article and Find Full Text PDF

Purpose: Plantar soft tissue properties affect foot biomechanics during movement. This study aims to explore the relationship between plantar pressure features and soft tissue stiffness through interpretable neural network model. The findings could inform orthotic insole design.

View Article and Find Full Text PDF

Motion analysis for the evaluation of dynamic spasticity during walking: A systematic scoping review.

Mult Scler Relat Disord

January 2025

Department of Neurology and Neurological Rehabilitation, Shanghai YangZhi Rehabilitation Hospital (Shanghai Sunshine Rehabilitation Center), School of Medicine, Tongji University, Shanghai 201619, China. Electronic address:

Background: Three-dimensional (3D) gait analysis has the potential to assess dynamic spasticity (DS). However, little is known about which parameters can be utilized for assessment.

Objective: To evaluate the application of 3D gait analysis in assessing DS during walking and to identify the most relevant parameters for clinical practice.

View Article and Find Full Text PDF

Phytic Acid-Induced Gradient Hydrogels for Highly Sensitive and Broad Range Pressure Sensing.

Adv Mater

January 2025

Laboratory of Advanced Theranostic Materials and Technology, Ningbo Institute of Materials Technology and Engineering, Chinese Academy of Sciences, Ningbo, 315201, P. R. China.

Ionic conductive hydrogels have emerged as an excellent option for constructing dielectric layers of interfacial iontronic sensors. Among these, gradient ionic hydrogels, due to the intrinsic gradient elastic modulus, can achieve a wide range of pressure responses. However, the fabrication of gradient hydrogels with optimal mechanical and sensing properties remains a challenge.

View Article and Find Full Text PDF

Lower limb biomechanics of chronic ankle instability (CAI) individuals has been widely investigated, but few have evaluated the internal foot mechanics in CAI. This study evaluated bone and soft tissue stress in CAI contrasted with copers and non-injured participants during a cutting task. Integrating scanned 3D foot shapes and free-form deformation, sixty-six personalized finite element foot models were developed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!