Rationale: High-temperature phosphoric acid digestion combined with continuous-flow isotope ratio mass spectrometry (CF-IRMS) is one of the standard methods for the determination of carbon and oxygen isotopes in carbonate. However, the routine purification protocol has not been tested by samples with nitrate contamination. Here we show that the standard protocol is not capable of measuring such samples with satisfactory accuracy and precision.
Methods: We tested the low-temperature (25°C) phosphoric acid digestion method combined with CF-IRMS for the measurement of natural carbonate samples with ~1% nitrate concentration and nitrate-doped carbonate standards. We also recorded the Raman spectra of the gases extracted from the nitrate-doped carbonate standards to identify the gas species that affect the results.
Results: The accuracy and precision of δ O values for nitrate-doped carbonate standards digested at 72°C are much poorer than routine measurements. For the measurements of doped standards reacted at 25°C, the accuracy and precision of the δ C and δ O values are indistinguishable from normal results. Raman spectra demonstrate that NO generated from high-temperature digestion is the possible interference species for oxygen isotope measurement.
Conclusions: Low-temperature digestion is a reliable method for the measurement of carbon and oxygen isotopes in nitrate-contaminated carbonate samples because the formation of NO can be largely reduced. This approach can be applied to some nitrate-rich speleothems, evaporites and other samples with nitrate contamination.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/rcm.8289 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!