Aims: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients.

Background: TWOS is a feared complication after implantation. It may lead to harmful IS. To date, specific strategies to reduce these events are lacking.

Methods: In this retrospective single-centre trial we analysed 146 S‑ICD patients who were implanted between 2010 and 2016. In all eligible consecutive patients (n = 139), follow-up of at least 6 weeks was studied. The incidence of TWOS/IS was analysed in patients receiving a 2 generation S‑ICD (Emblem-S-ICD) between 2014 and 2016 (Emblem). Their outcome was compared with a control group (SQ) treated with the SQ1010 device between 2010 and 2014, who were followed up for a maximum of 2 years. Furthermore, to test if the software update SMR8 reduces inappropriate shocks in the SQ1010-S-ICD population, the incidence of TWOS/IS was evaluated before and after update installation.

Results: Basic characteristics and indications for S‑ICD implantation were similar in both groups. However, the cumulative incidence of TWOS/IS was significantly decreased in Emblem vs. SQ (SQ: 15.4%, n = 14/91 vs. Emblem 4.2%, n = 2/48; p = 0.049). Furthermore, with regards to the SQ population we also observed a trend towards a significant reduction of TWOS/IS after installation of the software update SMR8 in 2014 (before update: 13.4%, n = 11/82 vs. after update: 4.6%, 3/65, p = 0.07).

Conclusion: 2 generation devices but probably also the SMR8 software update reduce the incidence of TWOS/IS in S‑ICD patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6288039PMC
http://dx.doi.org/10.1007/s12471-018-1160-yDOI Listing

Publication Analysis

Top Keywords

incidence twos/is
16
inappropriate shocks
12
software update
12
t‑wave oversensing
8
shocks subcutaneous
8
subcutaneous icd
8
s‑icd patients
8
update smr8
8
update
6
patients
5

Similar Publications

Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients.

Neth Heart J

December 2018

Division of Electrophysiology, Department of Cardiovascular Medicine, University of Muenster, Muenster, Germany.

Aims: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients.

Background: TWOS is a feared complication after implantation. It may lead to harmful IS.

View Article and Find Full Text PDF

Background: T-wave oversensing (TWOS) is a feared complication after subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation, potentially leading to inappropriate shocks (IS) with tremendous impact on quality of life.

Hypothesis: Postoperative ergometry facilitates primary and secondary prevention of TWOS or other potential causes of IS and optimizes S-ICD programming.

Methods: We analyze the impact of ergometry guided-programming (EMGP) on primary and secondary prevention of TWOS/IS in S-ICD patients, we screened 146 patients implanted in our center (2010-2016) for the incidence of TWOS/IS during postoperative ergometry.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!