Results of EMIT, Abuscreen RIA, and GC/MS tests for THC metabolites in a high volume random urinalysis program are compared. Samples were field tested by non-laboratory personnel with an EMIT system using a 100 ng/mL cutoff. Samples were then sent to the Army Forensic Toxicology Drug Testing Laboratory (WRAMC) at Fort Meade, Maryland, where they were tested by RIA (Abuscreen) using a statistical 100 ng/mL cutoff. Confirmations of all RIA positives were accomplished using a GC/MS procedure. EMIT and RIA results agreed for 91% of samples. Data indicated a 4% false positive rate and a 10% false negative rate for EMIT field testing. In a related study, results for samples which tested positive by RIA for THC metabolites using a statistical 100 ng/mL cutoff were compared with results by GC/MS utilizing a 20 ng/mL cutoff for the THCA metabolite. Presence of THCA metabolite was detected in 99.7% of RIA positive samples. No relationship between quantitations determined by the two tests was found.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/jat/10.5.178 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!