Objective: To compare the relaxation times acquired with synthetic MRI to those of multi-echo spin-echo sequences and to evaluate the usefulness of synthetic MRI in the clinical setting.

Methods: From January 2017 to May 2017, we included 51 patients with newly diagnosed breast cancer, who underwent additional synthetic MRI and multiecho spin echo (MESE) mapping sequences. Synthetic MRI technique uses a multiecho and multidelay acquisition method for the simultaneous quantification of physical properties such as and relaxation times and proton density image map. A radiologist with 9 years of experience in breast imaging drew region of interests manually along the tumor margins on two consecutive axial sections including the center of tumor mass and in the fat tissue of contralateral breast on both synthetic map and MESE map images.

Results: The mean relaxation time of the cancer was 84.75 ms (± 15.54) by synthetic MRI and 90.35 ms (± 19.22) by MESE mapping. The mean relaxation time of the fat was 129.22 ms (± 9.53) and 102.11 ms (± 13.9), respectively. Bland-Altman analysis showed mean difference of 8.4 ms for the breast cancer and a larger mean difference of 27.8 ms for the fat tissue. Spearman's correlation test showed that there was significant positive correlation between synthetic MRI and MESE sequences for the cancer ( = 0.713, < 0.001) and for the fat ( = 0.551, < 0.001). The positive estrogen receptor and low histologic grade were associated with little differences between two methods ( = 0.02  and = 0.043, respectively).

Conclusion: relaxation times of breast cancer acquired with synthetic MRI showed positive correlation with those of MESE mapping. Synthetic MRI could be useful for the evaluation of tissue characteristics by simultaneous acquisition of several quantitative physical properties.

Advances In Knowledge: Synthetic MRI is useful for the evaluation of relaxation times of the breast cancers.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6435064PMC
http://dx.doi.org/10.1259/bjr.20180479DOI Listing

Publication Analysis

Top Keywords

synthetic mri
40
breast cancer
16
relaxation times
16
relaxation time
12
mese mapping
12
mri
10
synthetic
10
multiecho spin
8
spin echo
8
acquired synthetic
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!