Background: Electronic devices for blood pressure (BP) measurements need to go through independent clinical validation as recommended by different authorities, both in general and in special populations such as pregnancy.

Objective: To evaluate the accuracy of the Omron Evolv (HEM-7600T-E) and the Omron M3 Comfort (HEM-7134-E) devices in pregnancy and pre-eclampsia according to the Universal Standard Validation Protocol.

Methods: Both devices, the Evolv and the M3 Comfort, measure BP at the brachial level using the oscillometric method. The study was performed according to the recently published protocol, the so-called "modified Advancement of Medical Instrumentation (AAMI)/British Hypertension Society (BHS)/European Society of Hypertension (ESH) protocol" or the "Universal Standard Protocol." Validation of each device included 45 pregnant women in the second and third gestational trimester of whom 15 had pre-eclampsia, 15 had gestational hypertension and 15 were normotensives. BP differences between the observer and the device BP values were classified into three categories (≤5, ≤10, and ≤15 mmHg) and the mean BP differences (test vs reference) and its SD were calculated.

Results: Both devices, the Evolv and the M3 Comfort, achieved a grade A/A in both pregnancy and pre-eclampsia. The mean difference (SD) between the mercury standard and the device BP values in pregnancy were: 1) for the Evolv of -0.7±2.3 mmHg for systolic blood pressure (SBP) and -0.1±1.8 mmHg for diastolic blood pressure (DBP); 2) for the M3 Comfort of -1.6±2.8 mmHg for SBP and -0.1±2.3 mmHg for DBP.

Conclusion: Both devices, the Evolv and the M3 Comfort, achieved a grade A/A for both SBP and DBP and fulfill the validation protocol criteria in pregnancy and pre-eclampsia. Consequently, these two devices can be recommended for home BP measurements in this specific population.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6124447PMC
http://dx.doi.org/10.2147/VHRM.S165524DOI Listing

Publication Analysis

Top Keywords

pregnancy pre-eclampsia
16
blood pressure
12
devices evolv
12
evolv comfort
12
accuracy omron
8
omron comfort
8
omron evolv
8
pressure measurements
8
universal standard
8
device values
8

Similar Publications

Unlabelled: Haemolysis, elevated liver enzymes and low platelets (HELLP) syndrome is a poorly understood, life-threatening multisystemic condition related to pregnancy with a rapid onset, typically observed in patients with severe pre-eclampsia. Various mechanisms may lead to diffuse endothelial damage associated with HELLP and possible brain involvement. A comprehensive review of PubMed, Embase and Cochrane databases was conducted to examine the clinical, laboratory and radiological features associated with postpartum HELLP syndrome, particularly its potential association with posterior reversible encephalopathy syndrome (PRES).

View Article and Find Full Text PDF

Effect of pre-conception serum vitamin D level on pregnancy outcomes in women undergoing in vitro fertilization with fresh embryo transfer: a retrospective analysis.

Reprod Biol Endocrinol

January 2025

Department of Obstetrics and Gynaecology, School of Clinical Medicine, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong Special Administrative Region, 6/F Professorial Block Queen Mary Hospital 102 Pok Fu Lam Road, Hong Kong, China.

Background: This study aims to investigate the association of pre-conception vitamin D levels on adverse pregnancy outcomes in women undergoing in vitro fertilization with fresh embryo transfer.

Methods: This was a retrospective cohort study using archived serum 25-hydroxyvitamin D measured in the pre-conception period before ovarian stimulation in patients undergoing in vitro fertilization with fresh autologous embryo transfer. A total of 306 women were included and adverse pregnancy outcomes in their resulting pregnancy were recorded.

View Article and Find Full Text PDF

Impact of interaction between individual genomes and preeclampsia on the severity of autism spectrum disorder symptoms.

Zhong Nan Da Xue Xue Bao Yi Xue Ban

August 2024

Hunan Key Laboratory of Medical Genetics, School of Life Sciences, Central South University, Changsha 410008.

Objectives: Autism spectrum disorder (ASD) is a complex neurodevelopmental disorder. Prior research suggests that genetic susceptibility and environmental exposures, such as maternal preeclampsia (PE) during pregnancy, play key roles in ASD pathogenesis. However, the specific effects of the interaction between genetic and environmental factors on ASD phenotype severity remain unclear.

View Article and Find Full Text PDF

Background: Aspirin has proved its efficacy in reducing the rate of preeclampsia in singleton pregnancy, however, there is discrepancy about the efficient dosage that should be used. While some societies recommend daily 75-81mg, others recommend higher dosage (160mg). This discrepancy is due to the lack of randomized controlled studies that compare these two dosages.

View Article and Find Full Text PDF

Regulation of macrophage polarization by metformin through inhibition of TLR4/NF-κB pathway to improve pre-eclampsia.

Placenta

January 2025

Department of Obstetrics and Gynecology, Affiliated Xiaoshan Hospital, Hangzhou Normal University, Hangzhou, Zhejiang, 311200, China. Electronic address:

Introduction: Pre-eclampsia (PE) is a pregnancy complication featuring hypertension and proteinuria. Metformin exerts clinically preventive effects on PE with an unspecified mechanism.

Methods: Placental tissues from PE patients and normal pregnant (NP) women were collected.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!